Comparing a head-mounted virtual reality perimeter and the Humphrey Field Analyzer for visual field testing in healthy and glaucoma patients.


ABSTRACT:

Purpose

To compare clinical visual field outputs in glaucoma and healthy patients returned by the Humphrey Field Analyzer (HFA) and virtual reality (Virtual Field, VF) perimetry.

Methods

One eye of 54 glaucoma patients and 41 healthy subjects was prospectively tested (three times each in random order) using the HFA and VF perimeters (24-2 test grids). We extracted and compared global indices (mean deviation [MD] and pattern standard deviation [PSD]), pointwise sensitivity (and calculated 'equivalent' sensitivity after accounting for differences in background luminance) and pointwise defects. Bland-Altman (mean difference [Mdiff ] and 95% limits of agreement [LoA]) and intraclass correlation analyses were performed.

Results

The VF test was shorter (by 76 s)