Advanced Vision Analyzer-Virtual Reality Perimeter: Device Validation, Functional Correlation and Comparison with Humphrey Field Analyzer.


ABSTRACT:

Purpose

To evaluate the Advanced Vision Analyzer (AVA; Elisar Vision Technology) and to compare pointwise threshold sensitivity and functional correlation of Elisar Standard Algorithm (ESA) with the Swedish Interactive Threshold Algorithm (SITA) of the Humphrey Field Analyzer (HFA; Carl Zeiss Meditec, Inc).

Design

Prospective, cross-sectional, observational case series.

Participants

One hundred sixty eyes (85 control participants, 75 glaucoma patients) for functional assessment, 15 eyes for test-retest variability (TRV), 107 eyes for blind spot trial (45 normal eyes, 62 glaucoma eyes) were recruited consecutively. A separate group of participants was chosen for each assessment.

Methods

All participants underwent ESA and SITA Standard 24-2 testing, and 1 eye of