High-Performance PMN-PT Single-Crystal-Based 1-3 Composite Transducer Integrated with a Biopsy Needle.


ABSTRACT: To address the need for high-resolution imaging in lung nodule detection and overcome the limitations of the shallow imaging depth associated with high-frequency ultrasound and the complex structure of lung tissue, we successfully integrated 50 MHz ultrasound transducers with 18-gauge biopsy needles. Featuring a miniaturized size of 0.6 × 0.5 × 0.5 mm3, the 50 MHz micromachined 1-3 composite transducer was tested to perform mechanical scanning of a nodule within a lung-tissue-mimicking phantom in vitro. The high-frequency transducer demonstrated the ability to achieve imaging with an axial resolution of 30 μm for measuring nodule edges. Moreover, the integrated biopsy needle prototype exhibited high accuracy (1.74% discrepancy) in estimating nodule area compared to actual dimensions in vitro. These results underscore the promising potential of biopsy-needle-integrated transducers in enhancing the accuracy of endoscopic ultrasound-guided fine needle aspiration biopsy (EUS-FNA) for clinical applications.

SUBMITTER: Kreager BC 

PROVIDER: S-EPMC10887013 | biostudies-literature | 2024 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

High-Performance PMN-PT Single-Crystal-Based 1-3 Composite Transducer Integrated with a Biopsy Needle.

Kreager Benjamin C BC   Wu Huaiyu H   Chang Wei-Yi WY   Moon Sunho S   Mitchell Josh J   Peng Chang C   Huang Chih-Chung CC   Muller Marie M   Tian Jian J   Jiang Xiaoning X  

Biosensors 20240131 2


To address the need for high-resolution imaging in lung nodule detection and overcome the limitations of the shallow imaging depth associated with high-frequency ultrasound and the complex structure of lung tissue, we successfully integrated 50 MHz ultrasound transducers with 18-gauge biopsy needles. Featuring a miniaturized size of 0.6 × 0.5 × 0.5 mm<sup>3</sup>, the 50 MHz micromachined 1-3 composite transducer was tested to perform mechanical scanning of a nodule within a lung-tissue-mimickin  ...[more]

Similar Datasets

| S-EPMC4772540 | biostudies-literature
| S-EPMC10160121 | biostudies-literature
| S-EPMC3543374 | biostudies-other
| S-EPMC5762724 | biostudies-literature
| S-EPMC9659027 | biostudies-literature
| S-EPMC6992459 | biostudies-literature
| S-EPMC3140242 | biostudies-other
| S-EPMC8901962 | biostudies-literature
| S-EPMC7744335 | biostudies-literature
| S-EPMC10204606 | biostudies-literature