ABSTRACT: Introduction
It is challenging for pregnant women with type 1 diabetes to maintain optimum glucose level to attain good neonatal outcomes. This study evaluated the efficacy of sensor-augmented insulin pump (SAP) with a predictive low-glucose suspend (PLGS) system in pregnant Japanese women with type 1 diabetes.Materials and methods
SAP with PLGS was used in 11 of the 22 women with type 1 diabetes who delivered between 2011 and 2021 at the two medical institutions in Japan. Glucose management, insulin delivery suspension time (IST) and neonatal outcomes were retrospectively studied.Results
In SAP with PLGS cases (n = 11), average glycated hemoglobin levels were < 6.5% throughout the pregnancy, and the time in range (TIR, 63-140 mg/dl) was > 70% in the second a