Optimization of the IntraVascular Oxygenator Catheter Using Angular Oscillation.
Ontology highlight
ABSTRACT: We demonstrate a methodology which both improves oxygen transport and reduces or eliminates bubble formation in a novel hyperbaric membrane oxygenator catheter model system. Angular oscillations were introduced to a bundle of hollow fiber membranes (HFMs) supplied with hyperbaric 100% oxygen at average gauge pressures up to 0.35 barg. Oscillating bundles enabled delivery of an oxygen flux of up to 400 mL min-1 m-2 in an aqueous solution, a doubling over a previous non-oscillating setup. Similarly, the addition of angular oscillations facilitated a five-fold reduction in pressure to achieve similar oxygen flux. The increased angular speed of oscillation improved flux, while the addition of angular micro-oscillation variations resulted in flux reductions of 7-20% compar