Ultra-fast searching assists in evaluating sub-ppm mass accuracy enhancement in U-HPLC/Orbitrap MS data.


ABSTRACT: A strategy, detailed methodology description and software are given with which the mass accuracy of U-HPLC-Orbitrap data (resolving power 50,000 FWHM) can be enhanced by an order of magnitude to sub-ppm levels. After mass accuracy enhancement all 211 reference masses have mass errors within 0.5 ppm; only 14 of these are outside the 0.2 ppm error margin. Further demonstration of mass accuracy enhancement is shown on a pre-concentrated urine sample in which evidence for 89 (342 ions) potential hydroxylated and glucuronated DHEA-metabolites is found. Although most DHEA metabolites have low-intensity mass signals, only 11 out of 342 are outside the ±1 ppm error envelop; 272 mass signals have errors below 0.5 ppm (142 below 0.2 ppm). The methodology consists of: (a) a multiple internal lock cor