ABSTRACT: Background
Septal penetration causes collimator-dependent differences in the heart-to-mediastinum (H/M) ratio in 123I-metaiodobenzylguanidine (MIBG) cardiac imaging. We investigated generally applicable methods to correct such differences.Methods and results
Four hours after 123I-MIBG injection, 40 patients underwent anterior chest imaging successively with medium-energy (ME) and various non-ME collimators. The H/M ratios obtained with the non-ME collimators before and after 123I-dual-window penetration correction were compared with the ME-derived standard values to determine patient-based conversion equations for empiric and combined corrections, respectively. A 123I point source was imaged with various collimators, and the central ratio, the ratio of count in a small central region of interest to count in a large one, was calculated. The method of predicting the conversion equations from the central ratios was determined. Correction using the patient-based conversion equations removed systematic underestimation of the H/M ratios obtained with the non-ME collimators, and combined correction depressed residual random errors to some degree. Point-source-based equations yielded results comparable to the patient-based equations.Conclusions
Empiric and combined corrections effectively reduce collimator-dependent differences in the H/M ratio. The conversion equations can be predicted from simple point-source imaging, which would allow to apply these corrections to data obtained with various collimators.
SUBMITTER: Inoue Y
PROVIDER: S-EPMC5629249 | biostudies-literature | 2017 Oct
REPOSITORIES: biostudies-literature
Inoue Yusuke Y Abe Yutaka Y Kikuchi Kei K Matsunaga Keiji K Masuda Ray R Nishiyama Kazutoshi K
Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology 20160601 5
<h4>Background</h4>Septal penetration causes collimator-dependent differences in the heart-to-mediastinum (H/M) ratio in <sup>123</sup>I-metaiodobenzylguanidine (MIBG) cardiac imaging. We investigated generally applicable methods to correct such differences.<h4>Methods and results</h4>Four hours after <sup>123</sup>I-MIBG injection, 40 patients underwent anterior chest imaging successively with medium-energy (ME) and various non-ME collimators. The H/M ratios obtained with the non-ME collimators ...[more]