A 30-Min Nucleic Acid Amplification Point-of-Care Test for Genital Chlamydia trachomatis Infection in Women: A Prospective, Multi-center Study of Diagnostic Accuracy.


ABSTRACT:

Background

Rapid Point-Of-Care Tests for Chlamydia trachomatis (CT) may reduce onward transmission and reproductive sexual health (RSH) sequelae by reducing turnaround times between diagnosis and treatment. The io® single module system (Atlas Genetics Ltd.) runs clinical samples through a nucleic acid amplification test (NAAT)-based CT cartridge, delivering results in 30min.

Methods

Prospective diagnostic accuracy study of the io® CT-assay in four UK Genito-Urinary Medicine (GUM)/RSH clinics on additional-to-routine self-collected vulvovaginal swabs. Samples were tested "fresh" within 10days of collection, or "frozen" at -80°C for later testing. Participant characteristics were collected to assess risk factors associated with CT infection.

Results

CT prevalence was 7.2% (51/709) overall. Sensitivity, specificity, positive and negative predictive values of the io® CT assay were, respectively, 96.1% (95% Confidence Interval (CI): 86.5-99.5), 97.7% (95%CI: 96.3-98.7), 76.6% (95%CI: 64.3-86.2) and 99.7% (95%CI: 98.9-100). The only risk factor associated with CT infection was being a sexual contact of an individual with CT.

Conclusions

The io® CT-assay is a 30-min, fully automated, high-performing NAAT currently CE-marked for CT diagnosis in women, making it a highly promising diagnostic to enable specific treatment, initiation of partner notification and appropriately intensive health promotion at the point of care.

SUBMITTER: Harding-Esch EM 

PROVIDER: S-EPMC5897918 | biostudies-literature | 2018 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

A 30-Min Nucleic Acid Amplification Point-of-Care Test for Genital Chlamydia trachomatis Infection in Women: A Prospective, Multi-center Study of Diagnostic Accuracy.

Harding-Esch E M EM   Cousins E C EC   Chow S-L C SC   Phillips L T LT   Hall C L CL   Cooper N N   Fuller S S SS   Nori A V AV   Patel R R   Thomas-William S S   Whitlock G G   Edwards S J E SJE   Green M M   Clarkson J J   Arlett B B   Dunbar J K JK   Lowndes C M CM   Sadiq S T ST  

EBioMedicine 20180110


<h4>Background</h4>Rapid Point-Of-Care Tests for Chlamydia trachomatis (CT) may reduce onward transmission and reproductive sexual health (RSH) sequelae by reducing turnaround times between diagnosis and treatment. The io® single module system (Atlas Genetics Ltd.) runs clinical samples through a nucleic acid amplification test (NAAT)-based CT cartridge, delivering results in 30min.<h4>Methods</h4>Prospective diagnostic accuracy study of the io® CT-assay in four UK Genito-Urinary Medicine (GUM)/  ...[more]

Similar Datasets

| S-EPMC2791188 | biostudies-literature
| S-EPMC5495747 | biostudies-literature
| S-EPMC6483758 | biostudies-literature
| S-EPMC5113142 | biostudies-literature
| S-EPMC11905427 | biostudies-literature
| S-EPMC2598672 | biostudies-literature
| S-EPMC262856 | biostudies-other
2013-03-01 | E-GEOD-41075 | biostudies-arrayexpress
2013-03-01 | GSE41075 | GEO
| S-EPMC529170 | biostudies-literature