Fabrication of Miniaturized Paper-Based Microfluidic Devices (MicroPADs).
Ontology highlight
ABSTRACT: Microfluidic paper-based analytical devices (microPADs) are emerging as cost-effective and portable platforms for point-of-care assays. A fundamental limitation of microPAD fabrication is the imprecise nature of most methods for patterning paper. The present work demonstrates that paper patterned via wax printing can be miniaturized by treating it with periodate to produce higher-resolution, high-fidelity microPADs. The optimal miniaturization parameters were determined by immersing microPADs in various concentrations of aqueous sodium periodate (NaIO4) for varying lengths of time. This treatment miniaturized microPADs by up to 80% in surface area, depending on the concentration of periodate and length of the reaction time. By immersing microPADs in 0.5-M NaIO4 for 48
SUBMITTER: Strong EB
PROVIDER: S-EPMC6327054 | biostudies-literature | 2019 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA