Rapid and Accurate Diagnosis of the Respiratory Disease Pertussis on a Point-of-Care Biochip.


ABSTRACT:

Background

Pertussis is a highly contagious respiratory disease caused by the bacterium Bordetella pertussis (B. pertussis). The infection is difficult to diagnose especially in underserved or resource-limited areas. We developed a low-cost and instrument-free diagnostic method for rapid and accurate detection of B. pertussis on a point-of-care (POC) testing device.

Methods

We developed a paper/polymer hybrid microfluidic biochip integrated with loop-mediated isothermal amplification (LAMP) method for the rapid and accurate detection of B. pertussis. This microfluidic approach was validated by testing 100 de-identified remnant clinical nasopharyngeal swabs and aspirates, which were confirmed to be either positive or negative for B. pertussis by a validated real-time PCR assay at the Children's Hospital Los Angeles.

Findings

The instrument-free detection results could be successfully read by the naked eye within 45 min with a limit of detection (LOD) of 5 DNA copies per well. Our optimized bacterial lysis protocol allowed the direct testing of clinical samples without any complicated sample processing/preparation (i.e. DNA extraction) or the use of any equipment (e.g. centrifuges). The validation of the microfluidic approach was accomplished by testing 100 clinical samples. High sensitivity (100%) and specificity (96%) with respect to real-time PCR were achieved.

Interpretation

This microfluidic biochip shows great potential for point-of-care disease diagnosis in various venues including schools and physician's offices, especially in low-resource settings in developing nations.

Funding

NIH/NIAID under award number R21AI107415, NIH RCMI Pilot Grant, the Philadelphia Foundation, the Medical Center of the Americas Foundation.

SUBMITTER: Dou M 

PROVIDER: S-EPMC6469871 | biostudies-literature | 2019 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Rapid and Accurate Diagnosis of the Respiratory Disease Pertussis on a Point-of-Care Biochip.

Dou Maowei M   Macias Natalie N   Shen Feng F   Bard Jennifer Dien JD   Domínguez Delfina C DC   Li Xiujun X  

EClinicalMedicine 20190201


<h4>Background</h4>Pertussis is a highly contagious respiratory disease caused by the bacterium <i>Bordetella pertussis</i> (<i>B. pertussis</i>). The infection is difficult to diagnose especially in underserved or resource-limited areas. We developed a low-cost and instrument-free diagnostic method for rapid and accurate detection of <i>B. pertussis</i> on a point-of-care (POC) testing device.<h4>Methods</h4>We developed a paper/polymer hybrid microfluidic biochip integrated with loop-mediated  ...[more]

Similar Datasets

| S-EPMC8831027 | biostudies-literature
| S-EPMC7721848 | biostudies-literature
| S-EPMC6019292 | biostudies-literature
| S-EPMC9033290 | biostudies-literature
| S-EPMC7076753 | biostudies-literature
| S-EPMC8217912 | biostudies-literature
| S-EPMC6598197 | biostudies-literature
| S-EPMC4761450 | biostudies-literature
| S-EPMC7233852 | biostudies-literature
| S-EPMC9494981 | biostudies-literature