Ontology highlight
ABSTRACT: Objectives
To assess the expansion pattern of coronary stents by using different balloon inflation times and pressures.Background
The selection of coronary stent size and its proper deployment is crucial in coronary artery interventions, having an impact on the success of the procedure and further therapy.Methods
Ten pairs of different stents were deployed under nominal pressure using sequential (5, 5, 10, and 10 seconds of repeated inflations, thus 30 seconds of summarized time) and continuous (30 seconds) deployment pattern. After each given time-point, intraluminal stent measurements were performed by optical coherence tomography (OCT) and intravascular ultrasound (IVUS).Results
Both in-stent diameters and cross-section areas (CSA) of paired stents measur
SUBMITTER: Skowronski J
PROVIDER: S-EPMC6739786 | biostudies-literature | 2019
REPOSITORIES: biostudies-literature