We evaluated the analytical performance of the fully automated cobas® 6500 urine work area and its automated components-cobas u 601 and cobas u 701.Design and methods
The study was conducted at three European centers using un-centrifuged surplus routine urine samples; all measurements were performed within 2 h of sample collection. Precision, sample carry-over, and method comparisons were evaluated per Clinical and Laboratory Standards Institute guidelines. Method comparisons: cobas u 601 versus Urisys 2400 and cobas u 411 urine test strips; and cobas u 701 versus KOVA® visual microscopy and iQ200 analyzer. Operability and functionality were assessed using questionnaires.Results
Precision of the entire cobas 6500 system was within predefined accept