Rapid detection of the New Delhi metallo-b-lactamase 1 (NDM-1) gene by loop-mediated isothermal amplification (LAMP).


ABSTRACT:

Background

New Delhi Metallo-b-lactamase (NDM-1) is an enzyme emerging around the world conferring resistance to a wide range of β-lactams agents and whose early detection is extremely important. We proposed to standardize the detection of the blaNDM-1 gene using the LOOP-mediated isothermal amplification technique (LAMP).

Methods

In all, 14 Gram-negative bacterial strains isolated from patients presenting pneumonia associated with mechanical ventilation were used for the blaNDM-1 standardization by LAMP. Klebsiella pneumoniae ATCC BAA-2473 and two clinical strains were used as a positive control. All results were compared to the reaction in polymerase chain reaction (PCR), considered gold standard for this detection.

Results

There was an excellent correlation between the two techniques employed, since all measured clinical strains were negative in both employed tests and two clinical, and a reference strains were positive.

Conclusions

The lamp technique seems to be an excellent option for the rapid detection of blaNDM-1. The amplification time is much shorter than other molecular techniques, the PCR machine is not necessary, it is easy of implementation and costs is low.

SUBMITTER: Moreira MG 

PROVIDER: S-EPMC6816916 | biostudies-literature | 2018 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Rapid detection of the New Delhi metallo-b-lactamase 1 (NDM-1) gene by loop-mediated isothermal amplification (LAMP).

Moreira Mirna Giselle MG   Barreto Lidia Miranda LM   Dos Santos Vera Lúcia VL   Monteiro Andrea Souza AS   Nobre Vandack V   Dos Santos Simone Gonçalves SG  

Journal of clinical laboratory analysis 20170928 4


<h4>Background</h4>New Delhi Metallo-b-lactamase (NDM-1) is an enzyme emerging around the world conferring resistance to a wide range of β-lactams agents and whose early detection is extremely important. We proposed to standardize the detection of the bla<sub>NDM</sub><sub>-1</sub> gene using the LOOP-mediated isothermal amplification technique (LAMP).<h4>Methods</h4>In all, 14 Gram-negative bacterial strains isolated from patients presenting pneumonia associated with mechanical ventilation were  ...[more]

Similar Datasets

| S-EPMC3347096 | biostudies-literature
| S-EPMC4627955 | biostudies-literature
| S-EPMC6113467 | biostudies-literature
| S-EPMC8650397 | biostudies-literature
| S-EPMC4444511 | biostudies-literature
| S-EPMC2238707 | biostudies-literature
| S-EPMC3000401 | biostudies-literature
| S-EPMC3212353 | biostudies-literature
| S-EPMC4033683 | biostudies-literature
| S-EPMC6709477 | biostudies-literature