In vitro characterization of PrismaLung+: a novel ECCO2R device.
Ontology highlight
ABSTRACT: BACKGROUND:Invasive mechanical ventilation is lifesaving in the setting of severe acute respiratory failure but can cause ventilation-induced lung injury. Advances in extracorporeal CO2 removal (ECCO2R) technologies may facilitate more protective lung ventilation in acute respiratory distress syndrome, and enable earlier weaning and/or avoid invasive mechanical ventilation entirely in chronic obstructive pulmonary disease exacerbations. We evaluated the in vitro CO2 removal capacity of the novel PrismaLung+ ECCO2R device compared with two existing gas exchangers. METHODS:The in vitro CO2 removal capacity of the PrismaLung+ (surface area 0.8 m2, Baxter) was compared with the PrismaLung (surface area 0.35 m2, Baxter) and A.L.ONE (surface area 1.35 m2, Eurosets) devices, using a closed-loop b
SUBMITTER: Hospach I
PROVIDER: S-EPMC7221037 | biostudies-literature | 2020 May
REPOSITORIES: biostudies-literature
ACCESS DATA