Inter- and intra-core laboratory variability in the quantitative coronary angiography analysis for drug-eluting stent treatment and follow up.


ABSTRACT:

Aim

To evaluate inter-core laboratory variability of quantitative coronary angiography (QCA) parameters in comparison with intra-core laboratory variability in a randomized controlled trial evaluating drug-eluting stents.

Methods

A total of 50 patients with 62 coronary lesions were analyzed by four analysis experts belonging to an Angiographic Core Laboratory (ACL: 1 expert) and a Cardiovascular Imaging Core Laboratory (CICL: 3 experts). QCA was based on the same standard operating procedure, but selections of projection and cine frames were at the discretion of each analyst. Inter- and intra-core laboratory variabilities were evaluated by accuracy, precision, Bland Altman analysis, and coefficient of variation.

Results

Pre-MLD (minimal lumen diameter) was significant