Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device.
Ontology highlight
ABSTRACT: Ambulatory blood pressure (BP) monitoring is recommended to improve the management of hypertension. Here, we investigated the accuracy of BP estimated using a wearable cuff-less device, InBodyWATCH, compared with BP measured using a manual sphygmomanometer. Thirty-five adults were enrolled (age 57.1 ± 17.9 years). The BP was estimated using InBodyWATCH with an individualized estimation based on a neural network model. Three paired sets of BPs from the two devices were compared using correlation analysis and Bland-Altman plots (n = 105 paired BP readings). The correlations for both systolic and diastolic BP (SBP and DBP) between the two devices were high (r = 0.964 and 0.939, both P < 0.001). The mean difference was 2.2 ± 6.1 mmHg for SBP and -0.2 ± 4.2 mmHg for DBP; these were not signific
SUBMITTER: Moon JH
PROVIDER: S-EPMC7642418 | biostudies-literature | 2020 Nov