Automatic treatment planning for VMAT-based total body irradiation using Eclipse scripting.
Ontology highlight
ABSTRACT: The purpose of this work is to establish an automated approach for a multiple isocenter volumetric arc therapy (VMAT)-based TBI treatment planning approach. Five anonymized full-body CT imaging sets were used. A script was developed to automate and standardize the treatment planning process using the Varian Eclipse v15.6 Scripting API. The script generates two treatment plans: a head-first VMAT-based plan for upper body coverage using four isocenters and a total of eight full arcs; and a feet-first AP/PA plan with three isocenters that covers the lower extremities of the patient. PTV was the entire body cropped 5 mm from the patient surface and extended 3 mm into the lungs and kidneys. Two plans were generated for each case: one to a total dose of 1200 cGy in 8 fractions and a second one t
SUBMITTER: Teruel JR
PROVIDER: S-EPMC7984467 | biostudies-literature | 2021 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA