Clinical Management and Pump Parameter Adjustment of the Control-IQ Closed-Loop Control System: Results from a 6-Month, Multicenter, Randomized Clinical Trial.
Ontology highlight
ABSTRACT: Background: Data are limited on the need for and benefits of pump setting optimization with automated insulin delivery. We examined clinical management of a closed-loop control (CLC) system and its relationship to glycemic outcomes. Materials and Methods: We analyzed personal parameter adjustments in 168 participants in a 6-month multicenter trial of CLC with Control-IQ versus sensor-augmented pump (SAP) therapy. Preset parameters (BR = basal rates, CF = correction factors, CR = carbohydrate ratios) were optimized at randomization, 2 and 13 weeks, for safety issues, participant concerns, or initiation by participants' usual diabetes care team. Time in range (TIR 70-180 mg/dL) was compared in the week before and after parameter changes. Results: In