Comparative evaluation of a dual-target real-time RT-PCR assay for COVID-19 diagnosis and assessment of performance in pooled saliva and nasopharyngeal swab samples.


ABSTRACT:

Objectives

Sensitive molecular diagnostic assays are essential for COVID-19 diagnosis. We evaluated the Hecin Scientific SARS-CoV-2 nucleic acid test kit, a dual-target real-time RT-PCR assay targeting the SARS-CoV-2 N and ORF1ab genes.

Methods

The Hecin test kit's diagnostic performance in detecting SARS-CoV-2 RNA was compared to the LightMix Modular SARS and Wuhan CoV E-gene kit (TIB Molbiol) and an in-house single-tube nested real-time RT-PCR using 296 clinical specimens, 11 proficiency testing samples, and 30 low-positive deep throat saliva and nasopharyngeal swab (NPS) samples pooled into negative samples in ratios of 1:5, 1:10, and 1:30.

Results

The limit-of-detection of the Hecin test kit was around 500 dC/mL for the N and ORF1ab targets. Sensitivity and specif