ABSTRACT: Objectives
We conducted a pilot study using an experimental study protocol to evaluate the measurement error of arterial pulse contour analysis-derived stroke volume due to improper transducer leveling during the passive leg raising test and the impact of such error on the determination of fluid responsiveness.Design
Prospective observational study.Setting
A medical-surgical ICU at a tertiary referral center in Kobe, Japan.Patients
Consecutive critically ill adult patients using the FloTrac system Version 4.0 (Edwards Lifesciences, Irvine, CA) for hemodynamic monitoring between September 1, 2018, and November 31, 2018.Interventions
None.Measurements and main results
Using 20 patients, we estimated the change in the zero-reference level of an