Regional ventilation distribution in healthy lungs: can reference values be established for electrical impedance tomography parameters?


ABSTRACT:

Background

Although electrical impedance tomography (EIT) is widely used for monitoring regional ventilation distribution, reference values have yet to be established for clinical use. The present study aimed to evaluate the feasibility of creating reference values for standard EIT parameters for potential clinical application.

Methods

A total of 75 participants with healthy lungs were included in this prospective study (male:female, 48:27; age, 34±14 years; height, 172±7 cm; weight, 73±12 kg). The subjects were examined during spontaneous breathing in the supine position. EIT measurements were performed at the level of the 4th intercostal space. Commonly used EIT-based parameters, including the center of ventilation (CoV), dorsal and most dorsal fractions of ventilation distribution (TVD and TVROI4 respectively), global inhomogeneity (GI) index, and standard deviation of regional ventilation delay index (RVDSD) were calculated.

Results

Following outlier detection, EIT data from 71 subjects were finally evaluated. The values of the evaluated parameters were: CoV, 48.7%±1.7%; TVD, 48.1%±5.4%; TVROI4, 7.1%±1.8%; GI, 0.49±0.04; and RVDSD, 7.0±2.0. The coefficients of variation for CoV and GI were low (0.03 and 0.07, respectively), but those for TVROI4 and RVDSD were comparatively high (0.26 and 0.28, respectively). None of the evaluated parameters showed a significant correlation with age. The GI index showed a weak but significant correlation with body mass index (R=0.29, P=0.01). The RVDSD was slightly higher in males than in females.

Conclusions

Our study indicated that CoV and GI were stable parameters with small coefficients of variation in participants with healthy lungs. The creation of EIT parameter reference values for setting treatment targets may be feasible.

SUBMITTER: Yang L 

PROVIDER: S-EPMC8246208 | biostudies-literature | 2021 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Regional ventilation distribution in healthy lungs: can reference values be established for electrical impedance tomography parameters?

Yang Lin L   Dai Meng M   Cao Xinsheng X   Möller Knut K   Dargvainis Mantas M   Frerichs Inéz I   Becher Tobias T   Fu Feng F   Zhao Zhanqi Z  

Annals of translational medicine 20210501 9


<h4>Background</h4>Although electrical impedance tomography (EIT) is widely used for monitoring regional ventilation distribution, reference values have yet to be established for clinical use. The present study aimed to evaluate the feasibility of creating reference values for standard EIT parameters for potential clinical application.<h4>Methods</h4>A total of 75 participants with healthy lungs were included in this prospective study (male:female, 48:27; age, 34±14 years; height, 172±7 cm; weig  ...[more]

Similar Datasets

| S-EPMC11484113 | biostudies-literature
| S-EPMC11251389 | biostudies-literature
| S-EPMC8327175 | biostudies-literature
| S-EPMC11863537 | biostudies-literature
| S-EPMC6999101 | biostudies-literature
| S-EPMC4705633 | biostudies-other
| S-EPMC7063910 | biostudies-literature
| S-EPMC6377686 | biostudies-literature
| S-EPMC8401348 | biostudies-literature
| S-EPMC10165791 | biostudies-literature