Comparison of 6 SARS-CoV-2 Molecular Methods and Correlation With the Cycle Threshold Distribution in Clinical Specimens.


ABSTRACT:

Background

The detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in patient samples is of critical importance in the management of patients and monitoring transmission in the population. However, data on the analytical performance characteristics for detection of SARS-CoV-2 in clinical specimens between individual targets within the same platform, and among different analytical platforms, are limited.

Methods

Here we evaluated the performance of 6 different sample-to-answer SARS-CoV-2 detection methods-Roche cobas 6800, Cepheid GeneXpert, Diasorin Simplexa, Luminex Aries emergency use authorization (EUA), Luminex Aries research use only (RUO), and bioMérieux BioFire-in clinical specimens with a range of viral loads.

Results

The positive percent