Hyperlipidemia Influences the Accuracy of Glucometer-Measured Blood Glucose Concentrations in Genetically Diverse Mice.


ABSTRACT:

Background

Glucometers are widely used in animal research due to simplicity and ease of utilization, but their accuracy in blood glucose assessment for hyperlipidemic mice is unknown.

Methods

Here, we compared blood glucose levels measured by a glucometer with plasma glucose levels measured by a standard enzymatic assay for 325 genetically diverse F2 mice derived from LP and BALB/c (BALB) Apoe-/- mice. Non-fasting glucose levels were measured before initiation of a Western diet and after 11 weeks on the diet.

Results

On chow diet, lab-measured plasma glucose levels were 279.5 ± 42.6 mg/dl (mean ± SD), while blood glucose values measured by glucometer were 138.7 ± 16.6 mg/dl. The two measures had no correlation (R2 = 0.006, p = 0.167). On the Western diet, plasma glucose levels rose to 351.1 ± 121.6 mg/dl, while glucometer-measured blood glucose fell to 128.7 ± 27.9 mg/dl. The two measures showed a moderate correlation (R2 = 0.111, p = 3.1E-9). Lab-measured plasma glucose showed strong positive correlations with plasma triglyceride and non-high-density lipoprotein cholesterol levels, while glucometer-measured blood glucose showed an inverse correlation with non-high-density lipoprotein levels on the chow diet.

Conclusions

Our results indicate that hyperlipidemia affects the accuracy of glucometers in measuring blood glucose levels of mice.

SUBMITTER: Shi LJ 

PROVIDER: S-EPMC8478864 | biostudies-literature | 2021 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Hyperlipidemia Influences the Accuracy of Glucometer-Measured Blood Glucose Concentrations in Genetically Diverse Mice.

Shi Lisa J LJ   Tang Xiwei X   He Jiang J   Shi Weibin W  

The American journal of the medical sciences 20210629 3


<h4>Background</h4>Glucometers are widely used in animal research due to simplicity and ease of utilization, but their accuracy in blood glucose assessment for hyperlipidemic mice is unknown.<h4>Methods</h4>Here, we compared blood glucose levels measured by a glucometer with plasma glucose levels measured by a standard enzymatic assay for 325 genetically diverse F2 mice derived from LP and BALB/c (BALB) Apoe<sup>-/-</sup> mice. Non-fasting glucose levels were measured before initiation of a West  ...[more]

Similar Datasets

| S-EPMC6199078 | biostudies-literature
| S-EPMC10421863 | biostudies-literature
| S-EPMC9520434 | biostudies-literature
| S-EPMC1087912 | biostudies-literature
| S-EPMC5127245 | biostudies-literature
| S-EPMC4724204 | biostudies-literature
| S-EPMC5627266 | biostudies-literature
| S-EPMC6940255 | biostudies-literature
| S-EPMC8317333 | biostudies-literature
| S-EPMC6565483 | biostudies-literature