The lifetime of a linac monitor unit ion chamber.


ABSTRACT: This study is the first to report the clinical lifetime of Varian Kapton sealed ion chambers as a retrospective review. The data have been analyzed using ion chamber gain values, daily quality assurance results, monthly quality assurance results, and delivered treatment field data were analyzed to comprehensively review trends. The data show the average lifetimes of the ion chambers from our institution, so other physicists can prepare for replacement. Additionally, we share our experience in performing quality assurance tests to calibrate and validate the radiation beam after ion chamber replacement.

SUBMITTER: Cetnar AJ 

PROVIDER: S-EPMC8664141 | biostudies-literature | 2021 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

The lifetime of a linac monitor unit ion chamber.

Cetnar Ashley J AJ   DiCostanzo Dominic J DJ  

Journal of applied clinical medical physics 20211111 12


This study is the first to report the clinical lifetime of Varian Kapton sealed ion chambers as a retrospective review. The data have been analyzed using ion chamber gain values, daily quality assurance results, monthly quality assurance results, and delivered treatment field data were analyzed to comprehensively review trends. The data show the average lifetimes of the ion chambers from our institution, so other physicists can prepare for replacement. Additionally, we share our experience in pe  ...[more]

Similar Datasets

| S-EPMC6843702 | biostudies-literature
| S-EPMC6537431 | biostudies-literature
| S-EPMC5153845 | biostudies-literature
| S-EPMC5892184 | biostudies-literature
| S-EPMC9561774 | biostudies-literature
| S-EPMC8436209 | biostudies-literature
| S-EPMC7010367 | biostudies-literature
| S-EPMC5875507 | biostudies-literature
| S-EPMC5721530 | biostudies-literature
| S-EPMC5912952 | biostudies-literature