Determination and Monitoring of Quality Parameters: A Detailed Study of Optical Elements of a Lens-Based Raman Spectrometer.
Ontology highlight
ABSTRACT: A lens-based Raman spectrometer is characterized by studying the optical elements in the optical path and we study the measure of aberration-diffraction effects. This is achieved by measuring the spectral resolution (SR) thus encompassing almost all optical elements of a spectrometer that are mostly responsible for such effects. An equation for SR is used to determine the quality factor Q which measures aberration/diffraction effects occurring in a spectrometer. We show how the quality factor changes with different spectrometer parameters such as grating groove density, the wavelength of excitation, pinhole width, charge-coupled device (CCD) pixel density, etc. This work provides an insight into the quality of a spectrometer and helps to monitor the performance of the spectrometer o
SUBMITTER: Mukherjee A
PROVIDER: S-EPMC8832553 | biostudies-literature | 2022 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA