Rotational Stability of the Clareon Monofocal Aspheric Hydrophobic Acrylic Intraocular Lens 6 Months After Implantation.


ABSTRACT:

Purpose

To determine the rotational stability of the Clareon® aspheric, monofocal, intraocular lens (IOL) up to 6 months after implantation.

Methods

This prospective, single-arm clinical study evaluated rotational stability of the Clareon IOL in a subset of subjects (n=141, 6 sites) that participated in an investigational device exemption trial for the Clareon IOL. The Clareon model (SY60CL) used in this subset was a non-toric IOL with toric axis markings to measure IOL rotation. All subjects (adults aged ≥22 years who required cataract extraction by phacoemulsification) received the Clareon IOL unilaterally. The position of the toric markings was captured using dilated retro-illumination slit-lamp photography and ocular anatomical landmarks. Post-operative rotational stability was assessed by an independent reading center. IOL rotation was defined as the difference between IOL axis of orientation on the day of surgery (≤1 hour after surgery) and each post-operative visit. Post-operative IOL-based rotational stability was evaluated at day 0 (day of surgery), day 1, week 1, month 1, and month 6 post-operatively.

Results

Compared with day 0, mean absolute IOL rotation was 1.85° on day 1 (n=127) and 2.27° at month 6 (n=124). Absolute IOL rotation ≤5° was observed in 95.3% of subjects on day 1 and 92.7% of subjects at month 6, compared with day 0. Between consecutive months 1 and 6 visits, mean absolute rotation was <1°; 100% of subjects had <10° rotation and 98.4% had ≤5°. The range of rotation on day 1 was 0° to 40.0° because of a subject with ocular trauma; when the trauma-outlier was removed, the mean absolute IOL rotation was 1.6° on day 1 (n=126) and 2.0° at month 6 (n=123).

Conclusion

These results support the high rotational stability of the Clareon monofocal IOL and serve as reference of the rotational stability of Clareon toric IOLs.

SUBMITTER: Walters TR 

PROVIDER: S-EPMC8858001 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature

altmetric image

Publications

Rotational Stability of the Clareon Monofocal Aspheric Hydrophobic Acrylic Intraocular Lens 6 Months After Implantation.

Walters Thomas R TR   Lehmann Robert R   Moyes Andrew A   French John W JW   Sreenivasan Vidhyapriya V   Modi Satish S SS  

Clinical ophthalmology (Auckland, N.Z.) 20220215


<h4>Purpose</h4>To determine the rotational stability of the Clareon<sup>®</sup> aspheric, monofocal, intraocular lens (IOL) up to 6 months after implantation.<h4>Methods</h4>This prospective, single-arm clinical study evaluated rotational stability of the Clareon IOL in a subset of subjects (n=141, 6 sites) that participated in an investigational device exemption trial for the Clareon IOL. The Clareon model (SY60CL) used in this subset was a non-toric IOL with toric axis markings to measure IOL  ...[more]

Similar Datasets

| S-EPMC10284131 | biostudies-literature
| S-EPMC7467255 | biostudies-literature
| S-EPMC9869798 | biostudies-literature
| S-EPMC8236945 | biostudies-literature
| S-EPMC10328104 | biostudies-literature
| S-EPMC3818402 | biostudies-literature
| S-EPMC8896947 | biostudies-literature
| S-EPMC9675354 | biostudies-literature
| S-EPMC11482183 | biostudies-literature
| S-EPMC9768760 | biostudies-literature