Impact of Design and Deployment Technique on the Hydrodynamic Resistance of Flow Diverters : An in Vitro Experimental Study.


ABSTRACT:

Purpose

Despite the high efficacy of flow diverters (FD) in treating sidewall intracranial aneurysms, failures are reported. One of the physical factors determining efficacy is the flow reducing capacity of the FD that is currently unknown to the operator. Our aim was to measure the flow reducing capacity expressed as the hydrodynamic resistance (HR), the metallic surface area (MSA) and pore density (PD) of two different FD designs and quantitatively investigate the impact of sizing and the deployment technique on these parameters.

Methods

Altogether 38 Pipeline (Medtronic) and P64 (Phenox) FD‑s were implanted in holder tubes by a neurointerventionist in nominally sized, oversized and longitudinally compressed or elongated manners. The tubes were placed in a flow model with the flow directed across the FD through a side hole on the tube. HR was expressed by the measured pressure drop as the function of the flow rate. Deployed length, MSA and PD were also measured and correlated with the HR.

Results

Both PD and MSA changed with varying deployment length, which correlates well with the change in HR. Oversizing the device by 1 mm in diameter has reduced the HR on average to one fifth of the original value for both manufacturers.

Conclusion

This study demonstrates experimentally that different FD designs have different flow diverting capacities (HR). Parameters are greatly influenced by radial sizing and longitudinal compression or elongation during implantation. Our results might be useful in procedure planning, predicting clinical outcome, and in patient-specific numerical flow simulations.

SUBMITTER: Gyurki D 

PROVIDER: S-EPMC8894302 | biostudies-literature | 2022 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Impact of Design and Deployment Technique on the Hydrodynamic Resistance of Flow Diverters : An in Vitro Experimental Study.

Gyürki Dániel D   Csippa Benjamin B   Paál György G   Szikora István I  

Clinical neuroradiology 20211022 1


<h4>Purpose</h4>Despite the high efficacy of flow diverters (FD) in treating sidewall intracranial aneurysms, failures are reported. One of the physical factors determining efficacy is the flow reducing capacity of the FD that is currently unknown to the operator. Our aim was to measure the flow reducing capacity expressed as the hydrodynamic resistance (HR), the metallic surface area (MSA) and pore density (PD) of two different FD designs and quantitatively investigate the impact of sizing and  ...[more]

Similar Datasets

| S-EPMC4054970 | biostudies-other
| S-EPMC6282594 | biostudies-literature
| S-EPMC7963647 | biostudies-literature
| S-EPMC7072062 | biostudies-literature
| S-EPMC8351456 | biostudies-literature
| S-EPMC5352494 | biostudies-literature
| S-EPMC6421840 | biostudies-literature
| S-EPMC7295809 | biostudies-literature
| S-EPMC7966473 | biostudies-literature
| S-EPMC8209672 | biostudies-literature