ABSTRACT: Purpose
Despite the high efficacy of flow diverters (FD) in treating sidewall intracranial aneurysms, failures are reported. One of the physical factors determining efficacy is the flow reducing capacity of the FD that is currently unknown to the operator. Our aim was to measure the flow reducing capacity expressed as the hydrodynamic resistance (HR), the metallic surface area (MSA) and pore density (PD) of two different FD designs and quantitatively investigate the impact of sizing and the deployment technique on these parameters.Methods
Altogether 38 Pipeline (Medtronic) and P64 (Phenox) FD‑s were implanted in holder tubes by a neurointerventionist in nominally sized, oversized and longitudinally compressed or elongated manners. The tubes were placed in a flow model with