ABSTRACT: Background
The purpose of the study was to dosimetrically compare multicatheter interstitial brachytherapy (MIBT) and stereotactic radiotherapy with CyberKnife (CK) for accelerated partial breast irradiation with special focus on dose to organs at risk (OARs).Materials and methods
Treatment plans of thirty-one patients treated with MIBT were selected and additional CK plans were created on the same CT images. The OARs included ipsilateral non-target and contralateral breast, ipsilateral and contralateral lung, skin, ribs, and heart for left sided cases. The fractionation was identical (4 × 6.25 Gy). Dose-volume parameters were calculated for both techniques and compared.Results
The D90 of the PTV for MIBT and CK were similar (102.4% vs. 103.6%, p = 0.0654), but in COIN the MIBT achieved lower value (0.75 vs. 0.91, p < 0.001). Regarding the V100 parameter of non-target breast CK performed slightly better than MIBT (V100: 1.1% vs. 1.6%), but for V90, V50 and V25 MIBT resulted in less dose. Every examined parameter of ipsilateral lung, skin, ribs and contralateral lung was significantly smaller for MIBT than for CK. Protection of the heart was slightly better with MIBT, but only the difference of D2cm3 was statistically significant (17.3% vs. 20.4%, p = 0.0311). There were no significant differences among the dose-volume parameters of the contralateral breast.Conclusion
The target volume can be properly irradiated by both techniques with high conformity and similar dose to the OARs. MIBT provides more advantageous plans than CK, except for dose conformity and the dosimetry of the heart and contralateral breast. More studies are needed to analyze whether these dosimetrical findings have clinical significance.
SUBMITTER: Herein A
PROVIDER: S-EPMC8989445 | biostudies-literature | 2022
REPOSITORIES: biostudies-literature
Herein András A Stelczer Gábor G Pesznyák Csilla C Fröhlich Georgina G Smanykó Viktor V Mészáros Norbert N Polgár Csaba C Takácsi-Nagy Zoltán Z Major Tibor T
Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology 20220322 1
<h4>Background</h4>The purpose of the study was to dosimetrically compare multicatheter interstitial brachytherapy (MIBT) and stereotactic radiotherapy with CyberKnife (CK) for accelerated partial breast irradiation with special focus on dose to organs at risk (OARs).<h4>Materials and methods</h4>Treatment plans of thirty-one patients treated with MIBT were selected and additional CK plans were created on the same CT images. The OARs included ipsilateral non-target and contralateral breast, ipsi ...[more]