Intercomparison of Radon Flux Monitors at Low and at High Radium Content Areas under Field Conditions.


ABSTRACT: Interlaboratory exercises are a good tool to compare the response of different systems to the same quantity and to identify possible inconsistencies between them. One of the main goals of the EMPIR 19ENV01 traceRadon project is to harmonize radon flux measurements based on different systems and methodologies. In the framework of the traceRadon Project, two radon flux intercomparison campaigns were carried out in October 2021 at high and at low radon source areas. Four institutions participated in the field intercomparison exercises with their own systems. Every system was based on a specific radon monitor (diffusion or pump mode) and an accumulation chamber (with manual or automatic opening). Radon fluxes were calculated by each participant using both exponential and linear fittings of the

SUBMITTER: Rabago D 

PROVIDER: S-EPMC8998188 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature