Improved simultaneous LET and dose measurements in proton therapy.


ABSTRACT: The objective of this study was to improve the precision of linear energy transfer (LET) measurements using [Formula: see text] optically stimulated luminescence detectors (OSLDs) in proton beams, and, with that, improve OSL dosimetry by correcting the readout for the LET-dependent ionization quenching. The OSLDs were irradiated in spot-scanning proton beams at different doses for fluence-averaged LET values in the (0.4-6.5) [Formula: see text] range (in water). A commercial automated OSL reader with a built-in beta source was used for the readouts, which enabled a reference irradiation and readout of each OSLD to establish individual corrections. Pulsed OSL was used to separately measure the blue (F-center) and UV ([Formula: see text]-center) emission bands of [Formula: see text] and the