ABSTRACT: Purpose
Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have emerged as a viable real-time indicator of tissue characteristics and the consequent durability of the lesions created. We investigated the impact of catheter-tissue contact force (CF) on LI behavior during pulmonary vein isolation (PVI).Methods
Forty-five consecutive patients of the CHARISMA registry undergoing de novo AF radiofrequency (RF) catheter ablation with a novel open-irrigated-tip catheter endowed with CF and LI measurement capabilities (Stablepoint™ catheter, Boston Scientific) were included.Results
A total of 2895 point-by-point RF applications were analyzed (RF delivery time (DT) = 8.7±4s, CF = 13 ±±8 g, LI drop = 23 ±±7 Ω). All PVs were successfully isol