Simulation Study of Radio Frequency Safety and the Optimal Size of a Single-Channel Surface Radio Frequency Coil for Mice at 9.4 T Magnetic Resonance Imaging.


ABSTRACT: The optimized size of a single-channel surface radio frequency (RF) coil for mouse body images in a 9.4 T magnetic resonance imaging (MRI) system was determined via electromagnetic-field analysis of the signal depth according to the size of a single-channel coil. The single-channel surface RF coils used in electromagnetic field simulations were configured to operate in transmission/reception mode at a frequency of 9.4 T-400 MHz. Computational analysis using the finite-difference time-domain method was used to assess the single-channel surface RF coil by comparing single-channel surface RF coils of varying sizes in terms of |B1|-, |B1+|-, |B1-|- and |E|-field distribution. RF safety for the prevention of burn injuries to small animals was assessed using an analysis of the specific absorption rate. A single-channel surface RF coil with a 20 mm diameter provided optimal B1-field distribution and RF safety, thus confirming that single-channel surface RF coils with ≥25 mm diameter could not provide typical B1-field distribution. A single-channel surface RF coil with a 20 mm diameter for mouse body imaging at 9.4 T MRI was recommended to preserve the characteristics of single-channel surface RF coils, and ensured that RF signals were applied correctly to the target point within RF safety guidelines.

SUBMITTER: Seo JH 

PROVIDER: S-EPMC9185248 | biostudies-literature | 2022 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Simulation Study of Radio Frequency Safety and the Optimal Size of a Single-Channel Surface Radio Frequency Coil for Mice at 9.4 T Magnetic Resonance Imaging.

Seo Jeung-Hoon JH   Ryu Yeunchul Y   Chung Jun-Young JY  

Sensors (Basel, Switzerland) 20220603 11


The optimized size of a single-channel surface radio frequency (RF) coil for mouse body images in a 9.4 T magnetic resonance imaging (MRI) system was determined via electromagnetic-field analysis of the signal depth according to the size of a single-channel coil. The single-channel surface RF coils used in electromagnetic field simulations were configured to operate in transmission/reception mode at a frequency of 9.4 T-400 MHz. Computational analysis using the finite-difference time-domain meth  ...[more]

Similar Datasets

| S-EPMC11251549 | biostudies-literature
| S-EPMC10167427 | biostudies-literature
| S-EPMC9536774 | biostudies-literature
| S-EPMC9694602 | biostudies-literature
| S-EPMC10137156 | biostudies-literature
| S-EPMC7002836 | biostudies-literature
| S-EPMC7408155 | biostudies-literature
| S-EPMC6232906 | biostudies-literature
| S-EPMC7889491 | biostudies-literature
| S-EPMC6109636 | biostudies-literature