Quantifying clinical severity of physics errors in high-dose rate prostate brachytherapy using simulations.


ABSTRACT:

Purpose

To quantitatively evaluate through automated simulations the clinical significance of potential high-dose rate (HDR) prostate brachytherapy (HDRPB) physics errors selected from our internal failure-modes and effect analysis (FMEA).

Methods and materials

A list of failure modes was compiled and scored independently by 8 brachytherapy physicists on a one-to-ten scale for severity (S), occurrence (O), and detectability (D), with risk priority number (RPN) = SxOxD. Variability of RPNs across observers (standard deviation/average) was calculated. Six idealized HDRPB plans were generated, and error simulations were performed: single (N = 1722) and systematic (N = 126) catheter shifts (craniocaudal; -1cm:1 cm); single catheter digitization errors (tip and connector needle-t

SUBMITTER: Nunez DA 

PROVIDER: S-EPMC9283911 | biostudies-literature | 2021 Sep-Oct