Quantifying clinical severity of physics errors in high-dose rate prostate brachytherapy using simulations.


ABSTRACT:

Purpose

To quantitatively evaluate through automated simulations the clinical significance of potential high-dose rate (HDR) prostate brachytherapy (HDRPB) physics errors selected from our internal failure-modes and effect analysis (FMEA).

Methods and materials

A list of failure modes was compiled and scored independently by 8 brachytherapy physicists on a one-to-ten scale for severity (S), occurrence (O), and detectability (D), with risk priority number (RPN) = SxOxD. Variability of RPNs across observers (standard deviation/average) was calculated. Six idealized HDRPB plans were generated, and error simulations were performed: single (N = 1722) and systematic (N = 126) catheter shifts (craniocaudal; -1cm:1 cm); single catheter digitization errors (tip and connector needle-tips displaced independently in random directions; 0.1 cm:0.5 cm; N = 44,318); and swaps (two catheters swapped during digitization or connection; N = 528). The deviations due to each error in prostate D90%, urethra D20%, and rectum D1cm3 were analyzed using two thresholds: 5-20% (possible clinical impact) and >20% (potentially reportable events).

Results

Twenty-nine relevant failure modes were described. Overall, RPNs ranged from 6 to 108 (average ± 1 standard deviation, 46 ± 23), with responder variability ranging from 19% to 184% (average 75% ± 30%). Potentially reportable events were observed in the simulations for systematic shifts >0.4 cm for prostate and digitization errors >0.3 cm for the urethra and >0.4 cm for rectum. Possible clinical impact was observed for catheter swaps (all organs), systematic shifts >0.2 cm for prostate and >0.4 cm for rectum, and digitization errors >0.2 cm for prostate and >0.1 cm for urethra and rectum.

Conclusions

A high variability in RPN scores was observed. Systematic simulations can provide insight in the severity scoring of multiple failure modes, supplementing typical FMEA approaches.

SUBMITTER: Nunez DA 

PROVIDER: S-EPMC9283911 | biostudies-literature | 2021 Sep-Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Purpose</h4>To quantitatively evaluate through automated simulations the clinical significance of potential high-dose rate (HDR) prostate brachytherapy (HDRPB) physics errors selected from our internal failure-modes and effect analysis (FMEA).<h4>Methods and materials</h4>A list of failure modes was compiled and scored independently by 8 brachytherapy physicists on a one-to-ten scale for severity (S), occurrence (O), and detectability (D), with risk priority number (RPN) = SxOxD. Variability  ...[more]

Similar Datasets

| S-EPMC6043748 | biostudies-literature
| S-EPMC4200189 | biostudies-literature
| S-EPMC6817536 | biostudies-literature
| S-EPMC6914687 | biostudies-literature
| S-EPMC7821293 | biostudies-literature
| S-EPMC10528916 | biostudies-literature
| S-EPMC7969770 | biostudies-literature
| S-EPMC7761604 | biostudies-literature
| S-EPMC8459608 | biostudies-literature
| S-EPMC9459480 | biostudies-literature