Performance evaluation of presepsin using a Sysmex HISCL-5000 analyzer and determination of reference interval.


ABSTRACT:

Background

Analytical evaluation of newly developed presepsin by a Sysmex HISCL-5000 (Sysmex, Japan) automated immune analyzer was performed.

Methods

For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of detection, method comparisons, and reference intervals were evaluated. Method comparisons were performed using a PATHFAST immune analyzer (LSI Medience Corporation, Japan).

Results

Precision using a 20x2x2 protocol for low (306 pg/mL) and high (1031 pg/mL) levels resulted in within-laboratory standard deviation (95% confidence interval [CI]) and coefficient of variation (CV) %, which were as follows: 15.3 (13.1-18.7), 5.5% and 47.7, (40.5-58.1), 6.4%, respectively. Linearity using patient samples and calibrators were measured from 201 to 16,177 and 188 to 30,000 pg/mL, respectively. The regression equation was y = -23.2 + 1.008x (SE = 162.4) for low levels and y = 779.9 + 1.006x (SE = 668) for high levels. Method comparison by Passing-Bablock analysis was as follows: y = -209.77 + 1.047x (Syx  = 335.3). The correlation coefficient (95% CI) was 0.869 (0.772-0.927) with statistical significance (p < 0.001). Reference intervals from 120 normal healthy subjects showed that 300 pg/mL was the cut off. Presepsin tended to show a higher value at higher ages and in males. Presepsin showed correlation with some parameters, and the correlation coefficient (p value) were as follows: hematocrit, 0.198 (0.03); eGFR (CKD-EPI), -0.240 (0.0129); MDRD-eGFR, -0.194 (0.048), respectively.

Conclusion

Presepsin measurement by HISCL-5000 showed reliable performance. Further clinical studies are required for the diagnosis and prognosis of sepsis.

SUBMITTER: Kang T 

PROVIDER: S-EPMC9459287 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance evaluation of presepsin using a Sysmex HISCL-5000 analyzer and determination of reference interval.

Kang Taewon T   Yoo Jeaeun J   Choi Hyunyu H   Lee Seungok S   Jekarl Dong Wook DW   Kim Yonggoo Y  

Journal of clinical laboratory analysis 20220723 9


<h4>Background</h4>Analytical evaluation of newly developed presepsin by a Sysmex HISCL-5000 (Sysmex, Japan) automated immune analyzer was performed.<h4>Methods</h4>For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of detection, method comparisons, and reference intervals were evaluated. Method comparisons were performed using a PATHFAST immune analyzer (LSI Medience Corporation, Japan).<h4>Results</h4>Pr  ...[more]

Similar Datasets

| S-EPMC9891513 | biostudies-literature
| S-EPMC6505781 | biostudies-literature
| S-EPMC6822009 | biostudies-literature
| S-EPMC10869398 | biostudies-literature
| S-EPMC6642327 | biostudies-literature
| S-EPMC8362000 | biostudies-literature
| S-EPMC2829900 | biostudies-literature
| S-EPMC6532615 | biostudies-literature
| S-EPMC8993620 | biostudies-literature
| S-EPMC9511180 | biostudies-literature