ABSTRACT: Purpose
To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice-encoded imaging direction and the reference sample ADC value.Methods
ADC acquisitions were performed monthly on an isotropic diffusion phantom over 1 year. Measurements of ADC %bias, coefficients of variation for short-/long-term repeatability and precision (CVST /CVLT and CVP ), and b-value dependency (Depb ) were calculated. The measurements were then assessed according to the Quantitative Imaging Biomarker Alliance (QI