Unknown

Dataset Information

0

Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Profile.


ABSTRACT:

Purpose

To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice-encoded imaging direction and the reference sample ADC value.

Methods

ADC acquisitions were performed monthly on an isotropic diffusion phantom over 1 year. Measurements of ADC %bias, coefficients of variation for short-/long-term repeatability and precision (CVST /CVLT and CVP ), and b-value dependency (Depb ) were calculated. The measurements were then assessed according to the Quantitative Imaging Biomarker Alliance (QIBA) Diffusion Profile specifications.

Results

The average of all measurements over the year was within Profile recommended ranges. This included when testing was performed in different imaging directions, and on samples that had different ADC reference values (0.4-1.1 μm2 /ms). Results in the axial plane for the central water vial included a bias of +0.05%, CVST /CVLT /CVP  = 0.1%/ 0.9%/0.4% and Depb  = 0.4%.

Conclusions

The technical performance of ADC on a radiotherapy dedicated MRI scanner over the course of 12 months was considered conformant to the QIBA Profile. Quantifying these metrics and factors that may affect the performance is essential in progressing the use of ADC clinically: ensuring that the observed change of ADC in a tissue is due to a physiological response and not measurement variability.

SUBMITTER: Carr ME 

PROVIDER: S-EPMC9543906 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Profile.

Carr Madeline E ME   Keenan Kathryn E KE   Rai Robba R   Boss Michael A MA   Metcalfe Peter P   Walker Amy A   Holloway Lois L  

Medical physics 20220411 7


<h4>Purpose</h4>To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice-encoded imaging direction and the reference sample ADC value.<h4>Methods</h4>ADC acquisitions were performed monthly on an isotropic diffusion phantom over 1 year. Measurements  ...[more]

Similar Datasets

| S-EPMC7244229 | biostudies-literature
| S-EPMC8480976 | biostudies-literature
| S-EPMC3800503 | biostudies-other
| S-EPMC5448876 | biostudies-other
| S-EPMC10923015 | biostudies-literature
| S-EPMC3904912 | biostudies-literature
| S-EPMC3115068 | biostudies-literature
| S-EPMC5911236 | biostudies-literature
| S-EPMC10243465 | biostudies-literature
| S-EPMC3745056 | biostudies-other