Bioprosthetic valve dysfunction and failure after TAVI in bicuspid aortic valve stenosis during one-year follow-up according to VARC-3.


ABSTRACT:

Background

Transcatheter aortic valve implantation (TAVI) in bicuspid aortic valve (BAV) stenosis has become more frequent in the last years. This may pose challenges for long-time valve durability. Therefore, we aimed to evaluate the prevalence of bioprosthetic valve dysfunction (BVD) with the newest-generation devices in BAV stenosis up to one-year follow-up (FU).

Methods

The primary endpoint was defined as the prevalence of BVD during the first procedural year according to Valve Academic Research Consortium (VARC)-3 criteria. Secondary endpoints were defined as failure in device success and clinical endpoints according to VARC-3.

Results

A total of 107 patients were included. Of these, 34 subjects (31.8%) met the criteria for BVD during a mean FU of 263 ± 180 days,