Biosensor Based on Peroxidase-Mimetic Nanozyme and Lactate Oxidase for Accurate L-Lactate Analysis in Beverages.


ABSTRACT: Precision analysis of the key biological metabolites such as L-lactate has great practical importance for many technological processes in food technology, including beverage production. Here we describe a new, highly selective, and sensitive biosensor for accurate L-lactate assay based on a combination of peroxidase-mimetic nanozymes with microbial lactate oxidase (LOx) immobilized onto the surface of a graphite-rod electrode (GE). The peroxidase-like nanozymes were synthesized using the debris of carbon microfibers (CFs) functionalized with hemin (H) and modified with gold nanoparticles (AuNPs) or platinum microparticles (PtMPs). The nanozyme formed with PtMPs as well as corresponding bioelectrodes based on it (LOx-CF-H-PtMPs/GE) is characterized by preferable catalytic and operational characteristics, so it was selected for the analysis of L-lactate content in real samples of grape must and red wine. The results of the L-lactate analysis obtained by the developed biosensors are highly correlated with a very selective spectrophotometric approach used as a reference. The developed biosensor, due to its high selectivity and sensitivity, is very prospective not only for the beverage industry and food technology, but also for clinical diagnostics and medicine, as well as in other applications where the accurate analysis of L-lactate is highly important.

SUBMITTER: Smutok O 

PROVIDER: S-EPMC9688602 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Biosensor Based on Peroxidase-Mimetic Nanozyme and Lactate Oxidase for Accurate L-Lactate Analysis in Beverages.

Smutok Oleh O   Kavetskyy Taras T   Prokopiv Tetiana T   Serkiz Roman R   Šauša Ondrej O   Novák Ivan I   Švajdlenková Helena H   Maťko Igor I   Gonchar Mykhailo M   Katz Evgeny E  

Biosensors 20221118 11


Precision analysis of the key biological metabolites such as L-lactate has great practical importance for many technological processes in food technology, including beverage production. Here we describe a new, highly selective, and sensitive biosensor for accurate L-lactate assay based on a combination of peroxidase-mimetic nanozymes with microbial lactate oxidase (LOx) immobilized onto the surface of a graphite-rod electrode (GE). The peroxidase-like nanozymes were synthesized using the debris  ...[more]

Similar Datasets

| S-EPMC8373162 | biostudies-literature
| S-EPMC8615781 | biostudies-literature
| S-EPMC5988403 | biostudies-literature
| S-EPMC9501957 | biostudies-literature
| S-EPMC6391499 | biostudies-literature
| S-EPMC9057007 | biostudies-literature
| S-EPMC6626002 | biostudies-literature
| S-EPMC11886749 | biostudies-literature
| S-EPMC9049964 | biostudies-literature
| S-EPMC5595127 | biostudies-literature