Evaluation of a new cemented highly cross-linked all-polyethylene cup: a prospective and randomised study assessing wear and fixation characteristics using radiostereometric analysis.


ABSTRACT:

Background and purpose

The aim of this prospective, randomised and controlled study was to evaluate the wear and fixation properties of a new cemented highly cross-linked all-polyethylene (HXLPE) cup in comparison with a conventional cemented ultra-high molecular weight polyethylene (ConvPE) cup using radiostereometric analysis (RSA).

Patients and methods

A total of 58 patients (58 hips) with primary osteoarthritis (OA) were enrolled in a randomised controlled trial to receive either a ConvPE cup (control) or HXLPE cup (intervention) with identical geometry. The subjects were randomised in a 1:1 ratio. The primary endpoint was proximal wear measured as femoral head penetration into the cup, secondary outcomes were 3D-wear and annual proximal wear from 1 to 5 years. Cup fixation was measured as movement of the cup in relation to the acetabular bone with proximal migration being the primary outcome measure, 3D-migration and change in inclination as secondary outcomes. The patients were followed for 5 years with RSA performed postoperatively, at 3, 12, 24, and 60 months.

Results

The HXLPE displayed a lower median proximal femoral head penetration compared to ConvPE, with a median difference at 2 years of -0.07 mm (95% CI, -0.10 to -0.04 mm), and -0.19 mm (95% CI, -0.27 to -0.15 mm) at 5 years. Annual proximal wear between 1 and 5 years was 0.03 mm/year for HXLPE and 0.06 mm/year for ConvPE (mean difference 0.05 mm, [95% CI, 0.03-0.07 mm]). Proximal migration, 3D migration and change in inclination was numerically slightly higher for HXLPE, albeit not statistically significant.

Conclusions

Compared to ConvPE, the HXLPE cup displayed significantly lower polyethylene wear. Cup migration was not statistically significant different.

Clinicaltrials.gov identifier

NCT04322799.

SUBMITTER: Otten V 

PROVIDER: S-EPMC9726741 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of a new cemented highly cross-linked all-polyethylene cup: a prospective and randomised study assessing wear and fixation characteristics using radiostereometric analysis.

Otten Volker V   Wästerlund Daniel D   Lindbjörn John J   Mertens Carl C   Mukka Sebastian S   Crnalic Sead S   Nilsson Kjell G KG  

Hip international : the journal of clinical and experimental research on hip pathology and therapy 20210210 6


<h4>Background and purpose</h4>The aim of this prospective, randomised and controlled study was to evaluate the wear and fixation properties of a new cemented highly cross-linked all-polyethylene (HXLPE) cup in comparison with a conventional cemented ultra-high molecular weight polyethylene (ConvPE) cup using radiostereometric analysis (RSA).<h4>Patients and methods</h4>A total of 58 patients (58 hips) with primary osteoarthritis (OA) were enrolled in a randomised controlled trial to receive eit  ...[more]

Similar Datasets

| S-EPMC4564778 | biostudies-literature
| S-EPMC4947731 | biostudies-literature
| S-EPMC8023918 | biostudies-literature
| S-EPMC11754158 | biostudies-literature
| S-EPMC10787387 | biostudies-literature
| S-EPMC9942043 | biostudies-literature
| S-EPMC8158183 | biostudies-literature
| S-EPMC9772923 | biostudies-literature
| S-EPMC7827971 | biostudies-literature
| S-EPMC8435938 | biostudies-literature