Non-invasive phrenic nerve stimulation to avoid ventilator-induced diaphragm dysfunction in critical care.


ABSTRACT:

Background

Diaphragm muscle atrophy during mechanical ventilation begins within 24 h and progresses rapidly with significant clinical consequences. Electrical stimulation of the phrenic nerves using invasive electrodes has shown promise in maintaining diaphragm condition by inducing intermittent diaphragm muscle contraction. However, the widespread application of these methods may be limited by their risks as well as the technical and environmental requirements of placement and care. Non-invasive stimulation would offer a valuable alternative method to maintain diaphragm health while overcoming these limitations.

Methods

We applied non-invasive electrical stimulation to the phrenic nerve in the neck in healthy volunteers. Respiratory pressure and flow, diaphragm electromyography and mechanomyography, and ultrasound visualization were used to assess the diaphragmatic response to stimulation. The electrode positions and stimulation parameters were systematically varied in order to investigate the influence of these parameters on the ability to induce diaphragm contraction with non-invasive stimulation.

Results

We demonstrate that non-invasive capture of the phrenic nerve is feasible using surface electrodes without the application of pressure, and characterize the stimulation parameters required to achieve therapeutic diaphragm contractions in healthy volunteers. We show that an optimal electrode position for phrenic nerve capture can be identified and that this position does not vary as head orientation is changed. The stimulation parameters required to produce a diaphragm response at this site are characterized and we show that burst stimulation above the activation threshold reliably produces diaphragm contractions sufficient to drive an inspired volume of over 600 ml, indicating the ability to produce significant diaphragmatic work using non-invasive stimulation.

Conclusion

This opens the possibility of non-invasive systems, requiring minimal specialist skills to set up, for maintaining diaphragm function in the intensive care setting.

SUBMITTER: Keogh C 

PROVIDER: S-EPMC9790411 | biostudies-literature | 2022 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Background</h4>Diaphragm muscle atrophy during mechanical ventilation begins within 24 h and progresses rapidly with significant clinical consequences. Electrical stimulation of the phrenic nerves using invasive electrodes has shown promise in maintaining diaphragm condition by inducing intermittent diaphragm muscle contraction. However, the widespread application of these methods may be limited by their risks as well as the technical and environmental requirements of placement and care. Non  ...[more]

Similar Datasets

| S-EPMC10728426 | biostudies-literature
| S-EPMC6348597 | biostudies-literature
| S-EPMC10118662 | biostudies-literature
| S-EPMC8500254 | biostudies-literature
| S-EPMC4365068 | biostudies-literature
| S-EPMC5913054 | biostudies-literature
| S-EPMC5699880 | biostudies-literature
| S-EPMC3947040 | biostudies-literature
| S-EPMC10816597 | biostudies-literature
| S-EPMC8301547 | biostudies-literature