ABSTRACT: Introduction
Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.Methods
Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using the novel monitor and a gold-standard foley manometer were compared. After anesthesia induction, a pneumoperitoneum was induced through a laparoscopic insufflator, and five randomly pre-defined pressures (between 5 and 25 mmHg) were achieved and simultaneously measured via both methods in each part