ABSTRACT: Introduction
Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.Methods
Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using the novel monitor and a gold-standard foley manometer were compared. After anesthesia induction, a pneumoperitoneum was induced through a laparoscopic insufflator, and five randomly pre-defined pressures (between 5 and 25 mmHg) were achieved and simultaneously measured via both methods in each participant. Measurements were compared using Bland-Altman analysis.Results
In total, 29 participants completed the study and provided 144 distinct pairs of pressure measurements that were analyzed. A positive correlation between the two methods was found (R2 = 0.93). There was good agreement between the methods, with a mean bias (95% CI) of -0.4 (-0.6, -0.1) mmHg and a standard deviation of 1.3 mmHg, which was statistically significant but of no clinical importance. The limits of agreement (where 95% of the differences are expected to fall) were -2.9 and 2.2 mmHg. The proportional error was statistically insignificant (p = 0.85), suggesting a constant agreement between the methods across the range of values tested. The percentage error was 10.7%.Conclusions
Continuous IAP measurements using the novel monitor performed well in the clinical setup of controlled intra-abdominal hypertension across the evaluated range of pressures. Further studies should expand the range to more pathological values.
SUBMITTER: Iacubovici L
PROVIDER: S-EPMC9967253 | biostudies-literature | 2023 Jan
REPOSITORIES: biostudies-literature
Iacubovici Liat L Karol Dana D Baar Yuval Y Beri Avi A Herzberg Haim H Zarour Shiri S Goren Or O Cohen Barak B
Life (Basel, Switzerland) 20230130 2
<h4>Introduction</h4>Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.<h4>Methods</h4>Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using ...[more]