Assessment of Intra-Abdominal Pressure with a Novel Continuous Bladder Pressure Monitor-A Clinical Validation Study.


ABSTRACT:

Introduction

Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.

Methods

Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using the novel monitor and a gold-standard foley manometer were compared. After anesthesia induction, a pneumoperitoneum was induced through a laparoscopic insufflator, and five randomly pre-defined pressures (between 5 and 25 mmHg) were achieved and simultaneously measured via both methods in each participant. Measurements were compared using Bland-Altman analysis.

Results

In total, 29 participants completed the study and provided 144 distinct pairs of pressure measurements that were analyzed. A positive correlation between the two methods was found (R2 = 0.93). There was good agreement between the methods, with a mean bias (95% CI) of -0.4 (-0.6, -0.1) mmHg and a standard deviation of 1.3 mmHg, which was statistically significant but of no clinical importance. The limits of agreement (where 95% of the differences are expected to fall) were -2.9 and 2.2 mmHg. The proportional error was statistically insignificant (p = 0.85), suggesting a constant agreement between the methods across the range of values tested. The percentage error was 10.7%.

Conclusions

Continuous IAP measurements using the novel monitor performed well in the clinical setup of controlled intra-abdominal hypertension across the evaluated range of pressures. Further studies should expand the range to more pathological values.

SUBMITTER: Iacubovici L 

PROVIDER: S-EPMC9967253 | biostudies-literature | 2023 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Assessment of Intra-Abdominal Pressure with a Novel Continuous Bladder Pressure Monitor-A Clinical Validation Study.

Iacubovici Liat L   Karol Dana D   Baar Yuval Y   Beri Avi A   Herzberg Haim H   Zarour Shiri S   Goren Or O   Cohen Barak B  

Life (Basel, Switzerland) 20230130 2


<h4>Introduction</h4>Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.<h4>Methods</h4>Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using  ...[more]

Similar Datasets

| S-EPMC9410406 | biostudies-literature
| S-EPMC10573363 | biostudies-literature
| S-EPMC8141480 | biostudies-literature
| S-EPMC9468118 | biostudies-literature
| S-EPMC3527153 | biostudies-literature
| S-EPMC8472078 | biostudies-literature
| S-EPMC4310251 | biostudies-literature
| S-EPMC3682312 | biostudies-literature
| 2228762 | ecrin-mdr-crc
| S-EPMC8521971 | biostudies-literature