Comparison of susceptibility-weighted angiography (SWAN) and T2 gradient-echo sequences for the detection of acute cerebral venous thrombosis.


ABSTRACT:

Objectives

To assess the diagnostic accuracy and lesion conspicuity of susceptibility-weighted angiography (SWAN) and T2* for the clot detection in acute cerebral venous thrombosis (CVT) by comparison with contrast-enhanced MR venography.

Methods

Venous thrombi detection and conspicuity were assessed by two readers for 18 venous segments on both T2*, SWAN source images, 2D SWAN reformats matching with T2*, and 3D SWAN images (SWAN-MinIP). Images obtained with the three reading techniques were systematically scored and compared to CE MRV findings, in a blinded fashion, per patient and per segment, and compared to each other.

Results

In 30 patients, 137 thrombosed venous segments were evaluated. The sensitivity of T2*, SWAN source images, 2D SWAN, and SWAN MinIP were, respectively, of 89.3%/82.1%, 82.1%, and 82.1% for dural sinus thrombosis and of 100%/100%/100%/96.6% for cortical venous thrombosis. There were significant differences in thrombus detection between T2* and SWAN: T2* versus SWAN source images and 2D SWAN (p = 0.04) and versus SWAN MinIP (p = 0.03). There were no significant differences between the three modalities of SWAN images. T2* was more sensitive than all SWAN images for both sigmoid sinus thrombosis and intracranial internal jugular vein thrombosis (p = 0.04). Inter-observer agreement was slightly superior with T2* (p < 0.05).

Conclusion

In this small cohort, SWAN sequence at 3T did not yield additional value for thrombus detection in acute CVT compared to T2*. This study highlights SWAN's greatest weakness both for diagnostic accuracy and lesion conspicuity compared to T2* for acute venous clot detection near the skull base.

SUBMITTER: Boukobza M 

PROVIDER: S-EPMC10034706 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of susceptibility-weighted angiography (SWAN) and T2 gradient-echo sequences for the detection of acute cerebral venous thrombosis.

Boukobza Monique M   Guinebert Sylvain S   Laissy Jean-Pierre JP  

The neuroradiology journal 20220618 2


<h4>Objectives</h4>To assess the diagnostic accuracy and lesion conspicuity of susceptibility-weighted angiography (SWAN) and T2* for the clot detection in acute cerebral venous thrombosis (CVT) by comparison with contrast-enhanced MR venography.<h4>Methods</h4>Venous thrombi detection and conspicuity were assessed by two readers for 18 venous segments on both T2*, SWAN source images, 2D SWAN reformats matching with T2*, and 3D SWAN images (SWAN-MinIP). Images obtained with the three reading tec  ...[more]

Similar Datasets

| S-EPMC4858464 | biostudies-literature
| S-EPMC9860026 | biostudies-literature
| S-EPMC9491445 | biostudies-literature
| S-EPMC10567825 | biostudies-literature
| S-EPMC8757924 | biostudies-literature
| S-EPMC3057399 | biostudies-literature
| S-EPMC8678192 | biostudies-literature
| S-EPMC7889861 | biostudies-literature
| S-EPMC8504525 | biostudies-literature
| S-EPMC8430322 | biostudies-literature