Unknown

Dataset Information

Performance of the VitaPCR rapid molecular test for SARS-CoV-2 screening at hospital admission.


ABSTRACT:

Objective

To evaluate the diagnostic accuracy of rapid VitaPCR™ (Credo) assay as screening test in emergency department (ED) patients prior to transfer or medical interventions.

Methods

In this prospective study 6642 oropharyngeal swabs from nonpreselected ED patients were tested for SARS-CoV-2 with (1) extraction-free VitaPCR and (2) extraction-based reference assays (Aptima®, cobas®, Xpert®Xpress).

Results

The median TAT of VitaPCR was 47 minutes (IQR: 38-59), while reference assays required 6.2 hours (IQR: 4.4-13.3). VitaPCR's sensitivity, specificity, PPV and NPV was 77.9%, 99.9%, 97.9%, and 98.9% in relation to Hologic Panther TMA; 78.3%, 99.8%, 96.4%, and 98.5% compared to Roche cobas6800 PCR; 71.2%, 99.2%, 94.9%, and 94.3% using Cepheid GeneXpert PCR as referen

SUBMITTER: Schuierer L 

PROVIDER: S-EPMC10154245 | biostudies-literature | 2023 May

REPOSITORIES: biostudies-literature