Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study.
Ontology highlight
ABSTRACT: In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless-and continual-BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled in a 12-week cardiac rehabilitation (CR) program (Neuchâtel, Switzerland). Daytime (9am-9pm) systolic (SBP) and diastolic (DBP) BP from 7-day averaged data from Aktiia monitor were compared to 1-day averaged BP data from ABPM. No significant differences were found between the Aktiia monitor and the ABPM for SBP (μ ± σ [95% confidence interval]: 1.6 ± 10.5 [-1.5, 4.6] mmHg, P = 0.306; correlation [R2]: 0.70; ± 10/ ± 15 mmHg agreements: 60%, 84%). Marginally non-significant bias was fou