Description, Feasibility, and Histological Assessment of the Vsling, a Novel Transcatheter Ventricular Repair Device.


ABSTRACT:

Background

Reshaping the dilated left ventricle using a surgically implanted papillary muscle sling has been shown to provide long-term improvement in cardiac function compared to annuloplasty alone in patients with systolic heart failure. A papillary muscle sling which can be implanted via a transcatheter approach has the potential to make this treatment more widely available to patients.

Methods

The Vsling transcatheter papillary muscle sling device was evaluated in a chronic animal model (sacrificed at 30 and 90 days), in a simulator, and in a human cadaver.

Results

The Vsling device was successfully implanted in 10 pigs, 6 simulator procedures, and 1 human cadaver. Procedure complexity and device usability were rated as reasonable or better by 6 interventional car