Validation of the AViTA BPM636 upper arm blood pressure monitor in adults and pregnant women according to the ANSI/AAMI/ISO 81060-2:2013.


ABSTRACT:

Objective

To evaluate the accuracy of the AViTA oscillometric upper arm home blood pressure (BP) monitor in adult and pregnant populations according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/ International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2013).

Methods

BP measurements on the upper arm were performed on 85 adult subjects and 46 pregnant subjects. The AViTA BPM636 and a standard mercury reference sphygmomanometer were applied and followed the same arm sequential BP measurement method. The universal cuff of the test device was used for arm circumference of 22-42 cm.

Results

For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 1.1 ± 5.49/2.9 ± 5.17 mmHg (systolic/diastolic) for adults; and -2.2 ± 5.93/1.5 ± 4.92 mmHg (systolic/diastolic) for pregnant women. For criterion 2, the SD of the averaged BP differences between the test device and reference BP per adult subject was 4.45/4.20 mmHg (systolic/diastolic) and per pregnant women was 4.66/3.96.

Conclusion

The AViTA BPM636 had passed the criteria of the ANSI/AAMI/ISO 81060-2:2013 protocol and can be recommended for home BP measurements in adults and pregnant populations.

SUBMITTER: Lee CN 

PROVIDER: S-EPMC10309106 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of the AViTA BPM636 upper arm blood pressure monitor in adults and pregnant women according to the ANSI/AAMI/ISO 81060-2:2013.

Lee Chien-Nan CN   Wu Cho-Kai CK   Huang I-Chih IC  

Blood pressure monitoring 20230405 4


<h4>Objective</h4>To evaluate the accuracy of the AViTA oscillometric upper arm home blood pressure (BP) monitor in adult and pregnant populations according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/ International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2013).<h4>Methods</h4>BP measurements on the upper arm were performed on 85 adult subjects and 46 pregnant subjects. The AViTA BPM636 and a sta  ...[more]

Similar Datasets

| S-EPMC11524622 | biostudies-literature
| S-EPMC9957726 | biostudies-literature
| S-EPMC10132455 | biostudies-literature
| S-EPMC9981319 | biostudies-literature
| S-EPMC7035100 | biostudies-literature
| S-EPMC9967253 | biostudies-literature
| S-EPMC7308245 | biostudies-literature
| S-EPMC11524633 | biostudies-literature
2023-06-03 | GSE233909 | GEO
| S-EPMC10200499 | biostudies-literature