Project description:The aim of this review is to summarize use of the functional lumen imaging probe (FLIP) for the evaluation and management of esophageal diseases.The FLIP utilizes high-resolution impedance planimetry to evaluate the functional geometry and cross-sectional area/pressure relationship (i.e. distensibility) of the esophageal sphincters and body. Recent studies have reported the potential utility of FLIP to evaluate esophageal diseases, such as achalasia, gastroesophageal reflux disease, and eosinophilic esophagitis.Esophageal distensibility measures provided by the FLIP will add to our understanding of the contribution of esophageal mechanical properties to esophageal symptoms and function. Application of the FLIP in clinical practice complements the diagnostic evaluation of esophageal function and may also be a valuable tool to help direct and target clinical management of esophageal diseases.
Project description:BackgroundThe functional lumen imaging probe (Endoflip™) is increasingly used for evaluation of patients with esophageal symptoms. To improve the interpretation of Endoflip™ in clinical practice, normative values with appropriate cut-off values are required.MethodsOriginal clinical studies describing Endoflip™ use for measurements of esophageal motility in healthy adults were considered. Meta-analysis was performed based on published values.ResultsA total of 17 articles were included in the systematic review, 15 of which were included in the meta-analysis, representing 154 unique subjects. At 40 ml distention, the 5th-95th and 10th-90th percentiles for esophagogastric junction distensibility index (EGJ-DI) were 1.96-10.95 mm2 /mmHg and 2.36-8.95 mm2 /mmHg, respectively. An EGJ-DI below 2 mm2 /mmHg was found in 5.4%, and below 3 mm2 /mmHg in 20.1% of healthy subjects. At 50 ml distention, the 5th-95th and 10th-90th percentiles for EGJ-DI are 2.86-10.66 mm2 /mmHg and 3.28-9.12 mm2 /mmHg, respectively (below 2 mm2 /mmHg: 0.6%, 3 mm2 /mmHg: 6.3%). The 5th-95th and 10th-90th percentiles for EGJ-DI at 60 ml distention were 3.06-8.14 mm2 /mmHg and 3.33-7.18 mm2 /mmHg, respectively (below 2 mm2 /mmHg: 0.0%, 3 mm2 /mmHg: 7%). A clear cut-off for lower values was identified while a large spread in values was observed for upper limits of normal for EGJ-DI for all filling volumes.ConclusionsGiven these observations, we recommend using a cut-off of 2 mm2 /mmHg for clinical practice, values below can be considered abnormal. Given that 5.4% of the healthy subjects will have an EGJ-DI below 2 mm2 /mmHg at 40 ml, we recommend using the 50 and 60 ml distention volumes. The clinical use of an upper limit for normality of EGJ-DI seems questionable.
Project description:Background/aimsThe mechanism via which supra-esophageal symptoms are generated is unclear. We assessed upper esophageal sphincter (UES) function in novel fashion using functional lumen imaging probe (FLIP) topography. We hypothesize that symptoms related to aspiration of esophageal contents may be associated with a more distensible UES.MethodsFLIP and reflux symptom index score data from patients undergoing diagnostic evaluation for an esophageal complaint over a 10-month period were analyzed retrospectively. UES distensibility on FLIP was studied at 40-70 mL volumes with in-depth analysis at 50 and 60 mL. Symptoms were compared between patients with low, middle, and high UES-distensibility index (UES-DI). Receiver-operating characteristic analysis was performed to determine associations between the UES-DI and individual reflux symptom index symptom item scores.ResultsOne hundred and eleven subjects were included. Overall, the associations between UES-DI and symptoms that could be related to supra-esophageal aspiration were strongest at the 50 mL FLIP volume. Choking item score was highest in the high UES-DI group (2.8) vs 1.4 (P < 0.001) in the middle UES-DI and 1.1 (P = 0.004) in the low UES-DI groups. Similarly, the cough item score was highest in the high UES-DI group (2.7) vs 1.5 (P = 0.009) and 0.9 (P = 0.002) groups.ConclusionA higher UES-DI measures defective barrier function which could may be the main pathophysiology that generates supra-esophageal symptoms.
Project description:Background/aimsThis study aimed to develop a diagnostic tool using machine learning to apply functional luminal imaging probe (FLIP) panometry data to determine the probability of esophagogastric junction (EGJ) obstruction as determined using the Chicago Classification version 4.0 (CCv4.0) and high-resolution manometry (HRM).MethodsFive hundred and fifty-seven adult patients that completed FLIP and HRM (with a conclusive CCv4.0 assessment of EGJ outflow) and 35 asymptomatic volunteers ("controls") were included. EGJ opening was evaluated with 16-cm FLIP performed during sedated endoscopy via EGJ-distensibility index and maximum EGJ diameter. HRM was classified according to the CCv4.0 as conclusive disorders of EGJ outflow or normal EGJ outflow (timed barium esophagram applied when required and available). The probability tool utilized Bayesian additive regression treesBART, which were evaluated using a leave-one-out approach and a holdout test set.ResultsPer HRM and CCv4.0, 243 patients had a conclusive disorder of EGJ outflow while 314 patients (and all 35 controls) had normal EGJ outflow. The model accuracy to predict EGJ obstruction (based on leave-one-out/holdout test set, respectively) was 89%/90%, with 87%/85% sensitivity, 92%/97% specificity, and an area under the receiver operating characteristic curve of 0.95/0.97. A free, open-source tool to calculate probability for EGJ obstruction using FLIP metrics is available at https://www.wklytics.com/nmgi/prob_flip.html.ConclusionsApplication of FLIP metrics utilizing a probabilistic approach incorporates the diagnostic confidence (or uncertainty) into the clinical interpretation of EGJ obstruction. This tool can provide clinical decision support during application of FLIP Panometry for evaluation of esophageal motility disorders.
Project description:Background/aimsThe functional lumen imaging probe (FLIP) device has been used to assess pyloric dysfunction in patients with gastroparesis. We aim to investigate whether varying FLIP catheter positions affect pyloric FLIP measurements.MethodsPatients undergoing endoscopy for chronic unexplained nausea and vomiting (CUNV) or gastroparesis were prospectively enrolled. FLIP balloon was adjusted for 3 positions within the pylorus: (1) proximal position, 75% of FLIP balloon in the duodenum and 25% in the antrum; (2) middle position, 50% in the duodenum and 50% in the antrum; and (3) distal position, 25% in the duodenum and 75% in the antrum. Pylorus cross-sectional area (CSA), intra-bag pressure (P), and distensibility indices (DI) were measured for 30, 40, and 50-mL balloon volumes. Fluoroscopic images were obtained to confirm FLIP balloon geometry. Data was analyzed separately using FLIP Analytic and customized MATLAB software.ResultsTwenty-two patients with CUNV (n = 4) and gastroparesis (n = 18) were enrolled. Pressures were significantly higher in the proximal position compared to the middle and distal positions. CSA measurements were significantly higher at the proximal and middle positions for 30-mL and 40-mL volume compared to the distal position values. DI values were significantly lower at the proximal positions for 40-mL and 50-mL distensions when compared to the middle and distal positions. Fluoroscopic images confirmed increased balloon bending when placed mostly in the duodenum.ConclusionsFLIP balloon position within the pylorus directly affects balloon geometry which significantly affects P, CSA, and DI measurements. Standardized pyloric FLIP protocols and balloon design adjustments are needed for the continued application of this technology to the pylorus.
Project description:Endoluminal functional lumen impedance planimetry (EndoFLIPTM) has become the gold standard to evaluate esophageal distensibility, although the study itself and its analysis present challenges. We propose here a new method to assess lower esophageal distension capacity that overcomes several limitations of prior approaches, including incomplete and corrupted EndoFLIPTM recordings. Esophageal distension capacity was evaluated with a 16-channel EndoFLIPTM in 10 controls and 14 patients with eosinophilic esophagitis (EoE). Controls were evaluated once. EoE patients were evaluated at baseline and after at least six weeks of treatment with orodispersible budesonide tablets, 1 mg bd. Balloon volumes were increased by 5 mL stepwise, either reaching a maximum volume of 60 mL or a maximum balloon pressure of 60 mmHg. Recordings were analyzed with a homemade R script. The mean esophageal diameter at 60 mL, D (60 mL), was calculated or extrapolated depending on whether the 60 mL volume was reached. By fitting a Michaelis-Menten curve across all measured diameters throughout all constant volume steps, the mean D (60 mL) was estimated. For control subjects, the mean ± SD value of D (60 mL) was 17.08 ± 1.69 mm, and for EoE patients at baseline, D (60 mL) was 14.51 ± 2.68 mm. After six weeks of treatment of EoE patients, D (60 mL) significantly increased to 16.22 ± 1.86 mm (paired Wilcoxon signed test: p = 0.0052), although the values for control subjects were not reached. The estimated mean esophageal diameter at 60 mL is a good proxy for esophageal distension capacity, which correlates with clinical outcomes in EoE. The method presented in this study overcomes difficulties encountered during the standard measurement protocol, allowing the analysis of recordings from incomplete and corrupted registries.
Project description:Background/aimsFunctional lumen imaging probe (FLIP) Panometry has demonstrated utility in the assessment of esophageal motility as a complement to existing methodologies like high-resolution manometry. However, as FLIP is typically performed with sedation during routine endoscopy, there is potential for impact of sedation agents on esophageal motility. We aim to examine the effects of conscious sedation with midazolam and fentanyl on FLIP Panometry metrics and classification.MethodsA cross-over study was conducted on 12 healthy, asymptomatic volunteers that completed FLIP while sedated with intravenous fentanyl and midazolam and while awake on a separate day. FLIP was performed in the same manner in both conditions with transoral placement of the FLIP and stepwise FLIP filling. During awake FLIP, subjects also rated the presence and intensity of esophageal perception.ResultsIn both experimental conditions, all subjects demonstrated normal motility. The esophagogastric junction distensibility index was lower (median [interquartile range]: 5.8 [5.15-6.85] vs 8.9 [7.68-9.38] mm2/mmHg; P = 0.025), and the FLIP pressure was higher (46.5 [38.125-52.5] vs 33 [26-36.8] mmHg; P = 0.010) in the sedated condition compared to the awake condition. Maximum esophagogastric junction diameter and body distensibility plateau were no different between conditions (P = 0.999 and P = 0.098, respectively). Perception of esophageal sensation during awake FLIP was reported in 7/12 (58%) subjects.ConclusionsWhile numeric differences in FLIP Panometry metrics were observed between sedated and awake FLIP in healthy subjects, these differences did not change the FLIP Panometry diagnosis. Sedated FLIP offers a well-tolerated method to assess esophageal motility during endoscopy.
Project description:BackgroundThe functional lumen imaging probe (FLIP) evaluates esophagogastric junction (EGJ) opening and esophageal contractility. Both post hoc and real-time analyses are possible, but reproducibility and reliability of analysis remain undefined. This study assesses inter- and intra-rater agreement of normative FLIP measurements among novice and experienced users.MethodsEight motility experts from different institutions independently evaluated de-identified video recordings from 27 asymptomatic healthy subjects using FLIP. Interpretation methods simulating a post-procedure and a live procedure setting were tested. Novice FLIP users (n = 3) received training prior to post-procedure interpretation. Experienced FLIP users (n = 5) interpreted using both methods. Users recorded maximum EGJ and distal esophageal body diameter, distensive pressure, and EGJ distensibility index (EGJ-DI), at balloon fill volumes of 50-, 60-, and 70 ml, as well as repetitive antegrade contractions (RACs). Inter- and intra-rater agreements of diameters, distensive pressure and EGJ-DI were assessed by intra-class correlation coefficient (ICC) and Pearson's correlation coefficient (PCC). Percentage agreement evaluated inter- and intra-rater reliability for RACs.Key resultsNovice and experienced users acquired normative FLIP metrics. Good-to-excellent inter- and intra-rater reliability were achieved for all variables at 60 ml balloon fill volumes. Median parameters at 60 ml balloon fill volume were as follows: EGJ-DI 5.5 mm2 /mmHg, maximum EGJ diameter 18.6 mm, distensive pressure at maximum EGJ diameter 48.1 mmHg, and distal esophageal body diameter 19.5 mm.Conclusions and inferencesNormative FLIP parameters can be reliably extracted from FLIP videos using both real-time and post hoc analyses, with high reliability between experienced and novice users.
Project description:INTRODUCTION:To compare the utility of the distensibility index (DI) on functional lumen imaging probe (FLIP) topography to other esophagogastric junction (EGJ) metrics in assessing treatment response in achalasia in the context of esophageal anatomy. METHODS:We prospectively evaluated 79 patients (at ages 17-81 years; 47% female patients) with achalasia during follow-up after pneumatic dilation, Heller myotomy, or per-oral endoscopic myotomy with timed barium esophagram, high-resolution impedance manometry, and FLIP. Anatomic deformities were identified based on consensus expert opinion. Patients were classified based on anatomy and EGJ opening to determine the association with radiographic outcome and Eckardt score (ES). RESULTS:Twenty-seven patients (34.1%) had an anatomic deformity-10 pseudodiverticula at myotomy, 7 epiphrenic diverticula, 5 sigmoid, and 5 sinktrap. A 5-minute column area of >5 cm was best associated with an ES of >3, with a sensitivity of 84% (P = 0.0013). Area under the curve for EGJ metrics in association with retention was as follows: DI, 0.90; maximal EGJ diameter, 0.76; integrated relaxation pressure, 0.64; and basal esophagogastric junction pressure, 0.53. Only FLIP metrics were associated with retention given normal anatomy (DI 2.4 vs 5.2 mm/mm Hg and maximal EGJ diameter 13.1 vs 16.6 mm in patients with and without retention, respectively; P values < 0.0001 and 0.002). Using a DI cutoff of <2.8 as abnormal, 40 of 45 patients with retention (P = 0.0001) and 23 of 25 patients with an ES of >3 (P = 0.02) had a low DI and/or anatomic deformity. With normal anatomy, 21 of 22 patients with retention had a low or borderline low DI. DISCUSSION:The FLIP DI is most useful metric for assessing the effect of achalasia treatment on EGJ opening. However, abnormal anatomy is an important mediator of outcome and treatment success will be modulated by anatomic defects that impede bolus emptying.