Technical note: Minimizing CIED artifacts on a 0.35 T MRI-Linac using deep learning.


ABSTRACT:

Background

Artifacts from implantable cardioverter defibrillators (ICDs) are a challenge to magnetic resonance imaging (MRI)-guided radiotherapy (MRgRT).

Purpose

This study tested an unsupervised generative adversarial network to mitigate ICD artifacts in balanced steady-state free precession (bSSFP) cine MRIs and improve image quality and tracking performance for MRgRT.

Methods

Fourteen healthy volunteers (Group A) were scanned on a 0.35 T MRI-Linac with and without an MR conditional ICD taped to their left pectoral to simulate an implanted ICD. bSSFP MRI data from 12 of the volunteers were used to train a CycleGAN model to reduce ICD artifacts. The data from the remaining two volunteers were used for testing. In addition, the dataset was reorganized three times using a Leave-One-Out scheme. Tracking metrics [Dice similarity coefficient (DSC), target registration error (TRE), and 95 percentile Hausdorff distance (95% HD)] were evaluated for whole-heart contours. Image quality metrics [normalized root mean square error (nRMSE), peak signal-to-noise ratio (PSNR), and multiscale structural similarity (MS-SSIM) scores] were evaluated. The technique was also tested qualitatively on three additional ICD datasets (Group B) including a patient with an implanted ICD.

Results

For the whole-heart contour with CycleGAN reconstruction: 1) Mean DSC rose from 0.910 to 0.935; 2) Mean TRE dropped from 4.488 to 2.877 mm; and 3) Mean 95% HD dropped from 10.236 to 7.700 mm. For the whole-body slice with CycleGAN reconstruction: 1) Mean nRMSE dropped from 0.644 to 0.420; 2) Mean MS-SSIM rose from 0.779 to 0.819; and 3) Mean PSNR rose from 18.744 to 22.368. The three Group B datasets evaluated qualitatively displayed a reduction in ICD artifacts in the heart.

Conclusion

CycleGAN-generated reconstructions significantly improved both tracking and image quality metrics when used to mitigate artifacts from ICDs.

SUBMITTER: Curcuru AN 

PROVIDER: S-EPMC10929994 | biostudies-literature | 2024 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Technical note: Minimizing CIED artifacts on a 0.35 T MRI-Linac using deep learning.

Curcuru Austen N AN   Yang Deshan D   An Hongyu H   Cuculich Phillip S PS   Robinson Clifford G CG   Gach H Michael HM  

Journal of applied clinical medical physics 20240218 3


<h4>Background</h4>Artifacts from implantable cardioverter defibrillators (ICDs) are a challenge to magnetic resonance imaging (MRI)-guided radiotherapy (MRgRT).<h4>Purpose</h4>This study tested an unsupervised generative adversarial network to mitigate ICD artifacts in balanced steady-state free precession (bSSFP) cine MRIs and improve image quality and tracking performance for MRgRT.<h4>Methods</h4>Fourteen healthy volunteers (Group A) were scanned on a 0.35 T MRI-Linac with and without an MR  ...[more]

Similar Datasets

| S-EPMC10800957 | biostudies-literature
| S-EPMC6800607 | biostudies-literature
| S-EPMC9310659 | biostudies-literature
| S-EPMC10069152 | biostudies-literature
| S-EPMC10860740 | biostudies-literature
| S-EPMC6443501 | biostudies-literature
| S-EPMC11575850 | biostudies-literature
| S-EPMC8906210 | biostudies-literature
| S-EPMC10521705 | biostudies-literature
| S-EPMC7098066 | biostudies-literature