ABSTRACT: Purpose
To measure the gradient system transfer function (GSTF) of an MR-Linac (Elekta Unity, Stockholm, Sweden) using an accessible phantom-based method and to apply trajectory corrections for UTE image reconstruction in the context of MR-guided radiotherapy of lung cancer.Methods
The first-order GSTF of a 1.5 T, split gradient Elekta Unity MR-Linac was measured using a thin-slice technique to characterize gradient system imperfections for each physical gradient axis (X, Y, Z). Repeatability measurements of the GSTF were performed 48 h apart. The GSTF was applied to trajectory correction in multi-echo UTE image reconstruction (TEs = 0.176, 1.85, 3.52 ms) to allow for UTE-Dixon inputs in the generation of synthetic CT. Images were acquired in an anthropomorphic phantom and