The CentriMag centrifugal blood pump as a benchmark for in vitro testing of hemocompatibility in implantable ventricular assist devices.


ABSTRACT: Implantable ventricular assist devices (VADs) have proven efficient in advanced heart failure patients as a bridge-to-transplant or destination therapy. However, VAD usage often leads to infection, bleeding, and thrombosis, side effects attributable to the damage to blood cells and plasma proteins. Measuring hemolysis alone does not provide sufficient information to understand total blood damage, and research exploring the impact of currently available pumps on a wider range of blood cell types and plasma proteins such as von Willebrand factor (vWF) is required to further our understanding of safer pump design. The extracorporeal CentriMag (Thoratec Corporation, Pleasanton, CA, USA) has a hemolysis profile within published standards of normalized index of hemolysis levels of less than 0.01?g/100?L at 100?mm?Hg but the effect on leukocytes, vWF multimers, and platelets is unknown. Here, the CentriMag was tested using bovine blood (n?=?15) under constant hemodynamic conditions in comparison with a static control for total blood cell counts, hemolysis, leukocyte death, vWF multimers, microparticles, platelet activation, and apoptosis. The CentriMag decreased the levels of healthy leukocytes (P?

SUBMITTER: Chan CH 

PROVIDER: S-EPMC4338790 | biostudies-literature | 2015 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

The CentriMag centrifugal blood pump as a benchmark for in vitro testing of hemocompatibility in implantable ventricular assist devices.

Chan Chris H H CH   Pieper Ina Laura IL   Hambly Rebecca R   Radley Gemma G   Jones Alyssa A   Friedmann Yasmin Y   Hawkins Karl M KM   Westaby Stephen S   Foster Graham G   Thornton Catherine A CA  

Artificial organs 20140728 2


Implantable ventricular assist devices (VADs) have proven efficient in advanced heart failure patients as a bridge-to-transplant or destination therapy. However, VAD usage often leads to infection, bleeding, and thrombosis, side effects attributable to the damage to blood cells and plasma proteins. Measuring hemolysis alone does not provide sufficient information to understand total blood damage, and research exploring the impact of currently available pumps on a wider range of blood cell types  ...[more]

Similar Datasets

| S-EPMC10927433 | biostudies-literature
| S-EPMC9225365 | biostudies-literature
| S-EPMC3343032 | biostudies-literature
| S-EPMC6601396 | biostudies-literature
| S-EPMC7416438 | biostudies-literature
| S-EPMC10024475 | biostudies-literature
| S-EPMC7301695 | biostudies-literature
| S-EPMC6662142 | biostudies-literature
| S-EPMC7386874 | biostudies-literature
| S-EPMC9132540 | biostudies-literature