Diagnostic accuracy of a pocket screening spirometer in diagnosing chronic obstructive pulmonary disease in general practice: a cross sectional validation study using tertiary care as a reference.


ABSTRACT:

Background

COPD-6™ is a lung function testing device for a rapid pre-spirometry testing to screen-out at-risk individuals not having COPD and indicating those at risk. The aim of this study was to validate COPD-6™ lung function testing (index test) in general practice in discriminating patients with COPD out of the population at risk - smokers/ex-smokers with no previous diagnosis of COPD, using measurements at tertiary care as reference standard.

Methods

Consecutive 227 subjects (115 women, 185 smokers/42 ex-smokers, ≥20 pack-years) with no previous diagnosis of COPD, aged 52.5 (SD 6.8) years from 26 general practitioners (GPs) were recruited, lung function tested with COPD-6™, referred to the tertiary institution for repeated COPD-6™ testing followed by spirometry with a bronchodilator (salbutamol), examination, and pulmonologist consultation for the diagnosis and severity of COPD.

Results

COPD was diagnosed in 43 subjects (18.9 %), with an AUC of 0.827 (95 % CI 0.769-0.875, P < 0.001) for the diagnosis of COPD when lung function was measured using COPD-6™ in GP's office with a specificity of 100 % (95 % CI, 97.95-100 %) but a very low sensitivity of 32.56 % (95 % CI, 20.49-47.48 %). Significant agreement for forced expiratory volume in 1 s measured at GP's office and at lung function lab was found (mean difference 0.01 L, p = 0.667) but not for other measured parameters (p < 0.001 for all).

Conclusions

Our study results point out that active case finding in a population at risk for COPD should be instituted (almost 20 % of undiagnosed COPD). Based on our results lung function testing with COPD-6™ can substitute spirometry testing in cases where it is not readily available to the patient/physician taken into account that the traditional FEV1/FEV6 cutoff value of <0.7 is not the only criterion for diagnosis and/or further referral.

Trial registration

ClinicalTrials.gov Identifier NCT01550679 Registered 28 September 2014, retrospectively registered.

SUBMITTER: Labor M 

PROVIDER: S-EPMC4992327 | biostudies-literature | 2016 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic accuracy of a pocket screening spirometer in diagnosing chronic obstructive pulmonary disease in general practice: a cross sectional validation study using tertiary care as a reference.

Labor Marina M   Vrbica Žarko Ž   Gudelj Ivan I   Labor Slavica S   Plavec Davor D  

BMC family practice 20160819 1


<h4>Background</h4>COPD-6™ is a lung function testing device for a rapid pre-spirometry testing to screen-out at-risk individuals not having COPD and indicating those at risk. The aim of this study was to validate COPD-6™ lung function testing (index test) in general practice in discriminating patients with COPD out of the population at risk - smokers/ex-smokers with no previous diagnosis of COPD, using measurements at tertiary care as reference standard.<h4>Methods</h4>Consecutive 227 subjects  ...[more]

Similar Datasets

| S-EPMC7575944 | biostudies-literature
| S-EPMC10423799 | biostudies-literature
| S-EPMC9741513 | biostudies-literature
| S-EPMC6852433 | biostudies-literature
| S-EPMC2632987 | biostudies-literature
| S-EPMC10989125 | biostudies-literature
| S-EPMC10589957 | biostudies-literature
| S-EPMC4772780 | biostudies-literature
| S-EPMC6545496 | biostudies-literature
| S-EPMC7677845 | biostudies-literature